This paper is only available as a PDF. To read, Please Download here.
A number of patients with silicone rubber implants performed by us and other surgeons
initially had excellent results; however, they returned with swelling and discomfort.
We studied 18 patients ranging in age from 16 years to 57 years who presented 8 to
78 months (average, 31.7 months) after silicone arthroplasty (four scaphoid, six lunate,
one scapholunate, four finger, two wrist, one trapezium, and one ulnar head for metacarpal
hemiarthroplasty). Erosive osteolysis was seen on x-ray films, with progressive destruction
evident in patients followed serially. None of the patients' conditions responded
to conservative care. The severity of the proliferative, inflammatory synovitis and
the foreign material in the multinucleated giant cells correlated with the interval
since arthroplasty. Implant surface analysis by scanning electron microscope and x-ray
spectrometer showed that silicone microparticles were the result of implant degeneration
and erosion. All joint cultures were negative. Silicone particulate synovitis and
destruction were arrested by the removal of the implant, a synovectomy, and curettage
of the lytic lesions at salvage (resection arthroplasty or arthrodesis.) Patients
who have had silicone arthroplasties should be followed indefinitely, at regular intervals,
by x-ray films and clinical examination.
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to Journal of Hand SurgeryAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- Review and analysis of silicone-rubber.metacarpophalangeal implants.J Bone Joint Surg [Am]. 1976; 58: 483-487
- Silastic scaphoid implants in osteoarthritis of the radioscaphoid joint.J Hand Surg. 1984; 9B: 177-180
- The arthritic great toe metatarso-phalangeal joint: A review of flexible silicone implant arthroplasty from two medical centers.Clin Orthop. 1981; 157: 64-69
- Metacarpophalangeal joint arthroplasty with silicone-dacron prostheses (Niebauer type): Six and a half years' experience.J Hand Surg. 1977; 2: 200-211
- The clinical experience with silicone-dacron metacarpophalangeal and interphalangeal joint prostheses.J Biomed Mater Res. 1973; 4: 137-163
- Kienbock's disease-update on silicone replacement arthroplasty.J Hand Surg. 1982; 7: 343-347
- Kienbock's disease: The role of silicone replacement arthroplasty.J Bone Joint Surg [Am]. 1977; 59: 899-908
- Silastic arthroplasty of the metacarpophalangeal joints in rheumatoid arthritis.J Bone Joint Surg [Am]. 1975; 57: 484-488
- The rheumatoid hand.Clin Orthop. 1984; 182: 150-159
- Metacarpophalangeal surgery in rheumatoid arthritis.Surg Clin North Am. 1969; 49: 823-831
- Treatment of Kienbock's disease using a silicone rubber implant.J Bone Joint Surg [Am]. 1976; 58: 373-376
- Use of hand-carved silicone rubber spacer for advanced Kienbock's disease.J Bone Joint Surg [Am]. 1981; 63: 1359-1370
- Wrist joint arthroplasty in rheumatoid arthritis: A comparison between the Meuli and Swanson prostheses.J Hand Surg. 1984; 9A: 171-176
- Silicone rubber implants for replacement of arthritic or destroyed joints in the hand.Surg Clin North Am. 1968; 48: 1113-1127
- Silicone rubber implants for the replacement of the carpal scaphoid and lunate bones.Orthop Clin North Am. 1970; 1: 299-309
- Flexible implant arthroplasty for arthritic finger joints: Rationale, technique, and results of treatment.J Bone Joint Surg [Am]. 1972; 54: 435-455
- Flexible implant arthroplasty for arthritic disabilities of the radiocarpal joint.A silicone rubber intramedullary stemmed flexible hinge implant for the wrist joint. Orthop Clin North Am. 1973; 4: 383-394
- Flexible implant arthroplasty in the hand.Clin Plast Surg. 1976; 3: 141-157
- Implant arthroplasty in the hand and upper extremity and its future.Surg Clin North Am. 1981; 61: 369-382
- Pathogenesis and pathomechanics of rheumatoid deformities in the hand and wrist.Orthop Clin North Am. 1973; 4: 1039-1056
- Flexible implant arthroplasty of the radiocarpal joint.Clin Orthop. 1984; 187: 94-106
- Flexible (silicone) implant arthroplasty in the small joints of extremities: Concepts, physical and biological considerations, experimental and clinical results.in: Rubin LR Biomaterials in reconstructive surgery. The CV Mosby Co, St. Louis1983: 595-623
- Treatment of selected patients with an ununited fracture of the proximal part of the scaphoid by excision of the fragment and insertion of a carved silicone rubber spacer.J Bone Joint Surg [Am]. 1984; 66: 510-517
- Look what you can make out of silicones.Chemtech. Sept, 1983; : 542-555
- Technology of silicones in biomedical applications.in: Rubin LR Biomaterials in reconstructive surgery. The CV Mosby Co, St. Louis1983: 73-90
- Silicone rubbers for medical applications.in: Williams DF Biocompatibility of clinical implant materials II. CRC Press Inc, Boca Raton1981: 79-98
- Fibrous capsule formation after subcutaneous implantation of synthetic materials in experimental animals.Plast Reconstr Surg. 1974; 54: 183-186
- Tissue reaction to pure and impregnated silastic.J Pediatr Surg. 1971; 6: 724-729
- Morphology of the connective tissue grown in response to implanted silicone rubber: A light and electron microscopic study.Surgery. 1974; 75: 631-637
- An experimental study of silicone as a soft tissue substitute.Plast Reconstr Surg. 1959; 24: 600-608
- Visceral response to subcutaneous and intraperitoneal injections of silicone in mice.Plast Reconstr Surg. 1967; 39: 402-410
- Reaction associated with a silicone rubber gel: An experimental study.J Biomed Mater Res. 1975; 9: 645-659
- Toxicologic studies, quality control, and efficacy of the silastic mammary prosthesis.Med Instrum. 1973; 7: 100-103
- The biostability of silicone rubbers, a polyamide, and a polyester.J Biomed Mater Res. 1976; 10: 123-143
- The effect of implantation on the physical properties of silicone rubber.J Biomed Mater Res. 1974; 8: 357-367
- 1984: F604-F678 (Philadelphia, ASTM: D412-D483 D624-D681; D813-D859 (1976);) 1984 Annual Book of ASTM Standards, V 9.01-9.02: Rubber, material and synthetic-general test methods: Carbon black.
- Letters to the editor.JAMA. 1977; 238: 939
- Letters to the editor.J Bone Joint Surg [Am]. 1983; 65: 280-281
- Long-term results of silicone implant arthroplasty.Implications of animal and human autopsy findings. JAMA. 1983; 250: 1195-1198
- Letters to the editor.JAMA. 1977; 238: 939
- Durability of silicone implants.An in-vivo study. Orthop Clin North Am. 1973; 4: 1097-1112
- Bone remodeling phenomena in flexible hand implants: Kappa Delta Awards.Orthop Review. 1982; 11: 129-130
- Foreign body reaction to silicone rubber: Complication of a finger joint implant.Clin Orthop. 1974; 98: 231-232
- Bleeding of silicone from bag-gel breast implants, and its chemical relation to fibrous capsule reaction.Plast Reconstr Surg. 1978; 61: 836-841
- Silicone lymphadenopathy and synovitis: Complications of silicone elastomer finger joint prostheses.JAMA. 1977; 237: 1463-1464
- Effect and fate of intra-articular silicone fluid.Clin Orthop. 1971; 77: 305-310
- Injectable fluid silicone therapy.Human morbidity and mortality. JAMA. 1976; 234: 308-309
- Complications of silicone implant surgery in the metacarpophalangeal joint.J Bone Joint Surg [Am]. 1975; 57: 991-994
- Pseudotumor of the arm associated with rupture of silicone-gel breast prostheses.J Bone Joint Surg [Am]. 1983; 65: 548-551
- Capsular contraction around silicone mammary prostheses.Ann Plast Surg. 1979; 2: 62-71
- Carpal collapse deformity following Innate resection for Kienböck's disease.Orth Trans. 1983; 7: 34
- Foreign-body reaction to silicone gel in axillary lymph nodes after an augmentation mammoplasty.Plast Reconstr Surg. 1978; 62: 381-384
- Complications following silicone injections for augmentation of the contours of the face.Plast Reconstr Surg. 1978; 61: 888-891
- The role of stress-enhanced reactivity in failure of orthopaedic implants.J Biomed Mater Res. 1973; 4: 401-418
- Destructive arthritis due to silicone: A foreign-body reaction.Radiology. 1983; 149: 69-72
- Foreign material in the capsules around breast prostheses and the cellular reaction to it.Br J Plast Surg. 1979; 32: 35-42
- Environmental fatigue testing of silastic finger joint prostheses.J Biomed Mater Res. 1972; 3: 15-24
- Reactive synovitis from particulate silastic.J Bone Joint Surg [Am]. 1982; 64: 581-585
- Fracture of silastic radial-head prosthesis.J Bone Joint Surg [Am]. 1981; 63: 459-460
- Silicone shard formation: A product of implant arthroplasty.J Foot Surg. 1984; 23: 362-365
- Silicone synovitis of the wrist.J Hand Surg. 1985; 10A: 47-61
- Stabilization of silastic implant arthroplasty at the trapezometacarpal joint.Clin Orthop. 1976; 121: 263-270
- Study and use of synthetic materials, such as silicones and teflon, as subcutaneous prostheses.Plast Reconstr Surg. 1960; 26: 264-279
- Biomedical evaluation of metacarpophalangeal joint prosthesis designs.J Hand Surg. 1979; 4: 508-521
- Tear energy of elastomers for maxillofacial applications.J Oral Rehabil. 1980; 7: 445-551
- Extensive localized bone resorption in the femur following total hip replacement.J Bone Joint Surg [Am]. 1976; 58: 612-618
- Flexible carbon fibre in late ligamentous reconstruction for instability of the knee.J Bone Joint Surg [Br]. 1984; 66: 196-200
- The foreign body reaction: A chronic inflammatory response.J Biomed Mater Res. 1974; 8: 199-211
- The ultrastructure of normal digital flexor tendon sheath and of the tissue formed around silicone and polyethylene implants in man.J Bone Joint Surg [Br]. 1977; 59: 206-212
- Induction of tendon sheath formation by the implantation of silicone tubes in rabbits.Br J Plast Surg. 1966; 19: 313-316
- Structure of capsules around silicone implants in hand surgery.Hand. 1983; 15: 47-52
- Histology of the capsules causing constrictive fibrosis around breast implants.Br J Plast Surg. 1977; 30: 324-329
- Vascularization and the gliding mechanism of free flexor-tendon grafts inserted by the silicone rod method.J Bone Joint Surg [Am]. 1974; 56: 473-482
- Experimental study of tissue response to ruptured gel-filled mammary prostheses.Plast Reconstr Surg. 1977; 59: 31-34
- Study of encapsulation of silicone rubber implants in animals: A foreignbody reaction.Plast Reconstr Surg. 1978; 62: 580-588
- The origin and nature of pseudosynovium appearing around implanted silastic rods: An experimental study.Hand. 1976; 8: 101-109
- Letters to the editor (reply).J Bone Joint Surg [Am]. 1983; 65: 281
- Gschwend N Debrunner HU Total hip prosthesis. The Williams & Wilkins Co, Baltimore1976: 205-239
- The role of soft tissue reconstruction in flexible implant arthroplasty of the metacarpophalangeal joint.Clin Orthop. 1979; 140: 178-183
- A rational postoperative management program for metacarpophalangeal joint implant arthroplasty.J Hand Surg. 1977; 2: 358-366
- Non-union of the scaphoid: Review of literature and current treatment.Orthop Rev. 1984; 13: 19-35
- The natural history of scaphoid nonunion.J Bone Joint Surg [Am]. 1984; 66: 504-509
- Guepar total knee prosthesis: Experience at the Vancouver General Hospital.J Bone Joint Surg [Br]. 1981; 63: 257-260
- An approach to Kienbock's disease: Triscaphe arthrodesis.J Hand Surg. 1985; 10A: 179-187
- Fractured femoral shaft through an osteolytic lesion resulting from the reaction to a prosthesis.J Bone Joint Surg [Br]. 1984; 66: 337-339
- Implants designed for finger joints.Scand J Plast Reconstr Surg. 1975; 9: 53-63
- Letters to the editor (reply).J Bone Joint Surg [Am]. 1983; 65: 281
- Synovial and osseous inflammation in failed silicone-rubber prostheses.J Bone Joint Surg [Am]. 1982; 64: 574-580
- Local and systemic effects of dimethylpolysiloxane fluid in mice.Plast Reconstr Surg. 1967; 39: 423-426
- Local and regional lymph node response to ruptured gel-filled mammary prostheses.Br J Plast Surg. 1978; 31: 349-352
- Three-dimensional imaging of the wrist.J Hand Surg. 1985; 10A: 32-39
- Giant bone cysts in rheumatoid arthritis.Orthop Rev. 1984; 13: 69-72
- Silicon identification in prosthesis-associated fibrous capsules.Science. 1978; 199: 437-479
Article info
Publication history
Accepted:
January 8,
1986
Received:
July 24,
1985
Identification
Copyright
© 1986 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.