Reactive synovitis after silicone arthroplasty

  • Clayton A. Peimer
    Correspondence
    Reprint requests: Clayton A. Peimer, M.D., Hand Center of Western N.Y., Millard Fillmore Hospital, 3 Gales Circle, Buffalo, NY 14209.
    Affiliations
    Departments of Orthopaedic Surgery, Mechanical Engineering, Anatomical Sciences, and Pathology, School of Medicine

    affiliated hospitals of the University at Buffalo, State University, N. Y.
    Search for articles by this author
  • John Medige
    Affiliations
    Departments of Orthopaedic Surgery, Mechanical Engineering, Anatomical Sciences, and Pathology, School of Medicine

    affiliated hospitals of the University at Buffalo, State University, N. Y.
    Search for articles by this author
  • Barry S. Eckert
    Affiliations
    Departments of Orthopaedic Surgery, Mechanical Engineering, Anatomical Sciences, and Pathology, School of Medicine

    affiliated hospitals of the University at Buffalo, State University, N. Y.
    Search for articles by this author
  • John R. Wright
    Affiliations
    Departments of Orthopaedic Surgery, Mechanical Engineering, Anatomical Sciences, and Pathology, School of Medicine

    affiliated hospitals of the University at Buffalo, State University, N. Y.
    Search for articles by this author
  • Craig S. Howard
    Affiliations
    Departments of Orthopaedic Surgery, Mechanical Engineering, Anatomical Sciences, and Pathology, School of Medicine

    affiliated hospitals of the University at Buffalo, State University, N. Y.
    Search for articles by this author
      This paper is only available as a PDF. To read, Please Download here.
      A number of patients with silicone rubber implants performed by us and other surgeons initially had excellent results; however, they returned with swelling and discomfort. We studied 18 patients ranging in age from 16 years to 57 years who presented 8 to 78 months (average, 31.7 months) after silicone arthroplasty (four scaphoid, six lunate, one scapholunate, four finger, two wrist, one trapezium, and one ulnar head for metacarpal hemiarthroplasty). Erosive osteolysis was seen on x-ray films, with progressive destruction evident in patients followed serially. None of the patients' conditions responded to conservative care. The severity of the proliferative, inflammatory synovitis and the foreign material in the multinucleated giant cells correlated with the interval since arthroplasty. Implant surface analysis by scanning electron microscope and x-ray spectrometer showed that silicone microparticles were the result of implant degeneration and erosion. All joint cultures were negative. Silicone particulate synovitis and destruction were arrested by the removal of the implant, a synovectomy, and curettage of the lytic lesions at salvage (resection arthroplasty or arthrodesis.) Patients who have had silicone arthroplasties should be followed indefinitely, at regular intervals, by x-ray films and clinical examination.
      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Hand Surgery
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Beckenbaugh RD
        • Dobyns JH
        • Linscheid RL
        • Bryan RS
        Review and analysis of silicone-rubber.metacarpophalangeal implants.
        J Bone Joint Surg [Am]. 1976; 58: 483-487
        • Bjomsson HA
        • Gestsson J
        • Ekelund L
        • Haffajee D
        Silastic scaphoid implants in osteoarthritis of the radioscaphoid joint.
        J Hand Surg. 1984; 9B: 177-180
        • Cracchiolo A
        • Swanson AB
        • deGroot Swanson G
        The arthritic great toe metatarso-phalangeal joint: A review of flexible silicone implant arthroplasty from two medical centers.
        Clin Orthop. 1981; 157: 64-69
        • Goldner JL
        • Gould JS
        • Urbaniak JR
        • McCollum DE
        Metacarpophalangeal joint arthroplasty with silicone-dacron prostheses (Niebauer type): Six and a half years' experience.
        J Hand Surg. 1977; 2: 200-211
        • Goldner JL
        • Urbaniak JR
        The clinical experience with silicone-dacron metacarpophalangeal and interphalangeal joint prostheses.
        J Biomed Mater Res. 1973; 4: 137-163
        • Lichtman DM
        • Alexander AH
        • Mack GR
        • Gunther SF
        Kienbock's disease-update on silicone replacement arthroplasty.
        J Hand Surg. 1982; 7: 343-347
        • Lichtman DM
        • Mack GR
        • MacDonald RI
        • Gunther SF
        • Wilson JN
        Kienbock's disease: The role of silicone replacement arthroplasty.
        J Bone Joint Surg [Am]. 1977; 59: 899-908
        • Mannerfelt L
        • Andersson K
        Silastic arthroplasty of the metacarpophalangeal joints in rheumatoid arthritis.
        J Bone Joint Surg [Am]. 1975; 57: 484-488
        • Nalebuff EA
        The rheumatoid hand.
        Clin Orthop. 1984; 182: 150-159
        • Nalebuff EA
        Metacarpophalangeal surgery in rheumatoid arthritis.
        Surg Clin North Am. 1969; 49: 823-831
        • Roca J
        • Beltran JE
        • Fairen MF
        • Alvarez A
        Treatment of Kienbock's disease using a silicone rubber implant.
        J Bone Joint Surg [Am]. 1976; 58: 373-376
        • Stark HH
        • Zemel NP
        • Ashworth CR
        Use of hand-carved silicone rubber spacer for advanced Kienbock's disease.
        J Bone Joint Surg [Am]. 1981; 63: 1359-1370
        • Sumners B
        • Hubbard MJS
        Wrist joint arthroplasty in rheumatoid arthritis: A comparison between the Meuli and Swanson prostheses.
        J Hand Surg. 1984; 9A: 171-176
        • Swanson AB
        Silicone rubber implants for replacement of arthritic or destroyed joints in the hand.
        Surg Clin North Am. 1968; 48: 1113-1127
        • Swanson AB
        Silicone rubber implants for the replacement of the carpal scaphoid and lunate bones.
        Orthop Clin North Am. 1970; 1: 299-309
        • Swanson AB
        Flexible implant arthroplasty for arthritic finger joints: Rationale, technique, and results of treatment.
        J Bone Joint Surg [Am]. 1972; 54: 435-455
        • Swanson AB
        Flexible implant arthroplasty for arthritic disabilities of the radiocarpal joint.
        A silicone rubber intramedullary stemmed flexible hinge implant for the wrist joint. Orthop Clin North Am. 1973; 4: 383-394
        • Swanson AB
        Flexible implant arthroplasty in the hand.
        Clin Plast Surg. 1976; 3: 141-157
        • Swanson AB
        Implant arthroplasty in the hand and upper extremity and its future.
        Surg Clin North Am. 1981; 61: 369-382
        • Swanson AB
        • deGroot Swanson G
        Pathogenesis and pathomechanics of rheumatoid deformities in the hand and wrist.
        Orthop Clin North Am. 1973; 4: 1039-1056
        • Swanson AB
        • deGroot Swanson G
        • Maupin BK
        Flexible implant arthroplasty of the radiocarpal joint.
        Clin Orthop. 1984; 187: 94-106
        • Swanson AB
        • deGroot Swanson G
        • Frisch EE
        Flexible (silicone) implant arthroplasty in the small joints of extremities: Concepts, physical and biological considerations, experimental and clinical results.
        in: Rubin LR Biomaterials in reconstructive surgery. The CV Mosby Co, St. Louis1983: 595-623
        • Zemel NP
        • Stark HH
        • Ashworth CR
        • Rickard TA
        • Anderson DR
        Treatment of selected patients with an ununited fracture of the proximal part of the scaphoid by excision of the fragment and insertion of a carved silicone rubber spacer.
        J Bone Joint Surg [Am]. 1984; 66: 510-517
        • Arkles LB
        Look what you can make out of silicones.
        Chemtech. Sept, 1983; : 542-555
        • Frisch EE
        Technology of silicones in biomedical applications.
        in: Rubin LR Biomaterials in reconstructive surgery. The CV Mosby Co, St. Louis1983: 73-90
        • van Noort R
        • Black MM
        Silicone rubbers for medical applications.
        in: Williams DF Biocompatibility of clinical implant materials II. CRC Press Inc, Boca Raton1981: 79-98
        • Inber G
        • Schwager RG
        • Guthrie RH
        • Gray GF
        Fibrous capsule formation after subcutaneous implantation of synthetic materials in experimental animals.
        Plast Reconstr Surg. 1974; 54: 183-186
        • Irving IM
        • Castilla P
        • Hall EG
        • Rickham PP
        Tissue reaction to pure and impregnated silastic.
        J Pediatr Surg. 1971; 6: 724-729
        • Jauregui PH
        • Gascia CE
        • Angulo AG
        Morphology of the connective tissue grown in response to implanted silicone rubber: A light and electron microscopic study.
        Surgery. 1974; 75: 631-637
        • Marzoni FA
        • Upchurch SE
        • Lambert CJ
        An experimental study of silicone as a soft tissue substitute.
        Plast Reconstr Surg. 1959; 24: 600-608
        • Rees TD
        • Ballantyne Jr, DL
        • Seidman I
        • Hawthorne GA
        Visceral response to subcutaneous and intraperitoneal injections of silicone in mice.
        Plast Reconstr Surg. 1967; 39: 402-410
        • Rigdon RH
        • Dricks A
        Reaction associated with a silicone rubber gel: An experimental study.
        J Biomed Mater Res. 1975; 9: 645-659
        • Robertson G
        • Braley S
        Toxicologic studies, quality control, and efficacy of the silastic mammary prosthesis.
        Med Instrum. 1973; 7: 100-103
        • Roggendorf E
        The biostability of silicone rubbers, a polyamide, and a polyester.
        J Biomed Mater Res. 1976; 10: 123-143
        • Swanson JW
        • Lebeau JE
        The effect of implantation on the physical properties of silicone rubber.
        J Biomed Mater Res. 1974; 8: 357-367
        • American Society for Testing and Materials
        1984 Annual Book of ASTM Standards, V 9.01-9.02: Rubber, material and synthetic-general test methods: Carbon black. 1984: F604-F678 (Philadelphia, ASTM: D412-D483 D624-D681; D813-D859 (1976);)
        • Nalbandian RM
        Letters to the editor.
        JAMA. 1977; 238: 939
        • Nalbandian RM
        Letters to the editor.
        J Bone Joint Surg [Am]. 1983; 65: 280-281
        • Nalbandian RM
        • Swanson AB
        • Maupin BK
        Long-term results of silicone implant arthroplasty.
        Implications of animal and human autopsy findings. JAMA. 1983; 250: 1195-1198
        • Swanson AB
        Letters to the editor.
        JAMA. 1977; 238: 939
        • Swanson AB
        • Meester WD
        • deGroot Swanson G
        • Rangaswamy L
        • Schut GED
        Durability of silicone implants.
        An in-vivo study. Orthop Clin North Am. 1973; 4: 1097-1112
        • Swanson AB
        • Poitevin LA
        • deGroot Swanson G
        Bone remodeling phenomena in flexible hand implants: Kappa Delta Awards.
        Orthop Review. 1982; 11: 129-130
        • Aptekar RG
        • Davie JM
        • Cattell HS
        Foreign body reaction to silicone rubber: Complication of a finger joint implant.
        Clin Orthop. 1974; 98: 231-232
        • Barker DE
        • Retsky MI
        • Schultz S
        Bleeding of silicone from bag-gel breast implants, and its chemical relation to fibrous capsule reaction.
        Plast Reconstr Surg. 1978; 61: 836-841
        • Christie AJ
        • Weinberger KA
        • Dietrich M
        Silicone lymphadenopathy and synovitis: Complications of silicone elastomer finger joint prostheses.
        JAMA. 1977; 237: 1463-1464
        • Donahue WC
        • Nosanchuk JS
        • Kaufer H
        Effect and fate of intra-articular silicone fluid.
        Clin Orthop. 1971; 77: 305-310
        • Ellenbogen R
        • Rubin L
        Injectable fluid silicone therapy.
        Human morbidity and mortality. JAMA. 1976; 234: 308-309
        • Ferlic DC
        • Clayton ML
        • Holloway M
        Complications of silicone implant surgery in the metacarpophalangeal joint.
        J Bone Joint Surg [Am]. 1975; 57: 991-994
        • Foster WC
        • Springfield DS
        • Brown KLB
        Pseudotumor of the arm associated with rupture of silicone-gel breast prostheses.
        J Bone Joint Surg [Am]. 1983; 65: 548-551
        • Gayou R
        • Rudolph R
        Capsular contraction around silicone mammary prostheses.
        Ann Plast Surg. 1979; 2: 62-71
        • Hastings III, H
        • Thornberry RL
        • Kleinman WB
        • Strickland JW
        • Steichen JB
        Carpal collapse deformity following Innate resection for Kienböck's disease.
        Orth Trans. 1983; 7: 34
        • Hausner RJ
        • Schoen FJ
        • Pierson KK
        Foreign-body reaction to silicone gel in axillary lymph nodes after an augmentation mammoplasty.
        Plast Reconstr Surg. 1978; 62: 381-384
        • Pearl RM
        • Laub DR
        • Kaplan EN
        Complications following silicone injections for augmentation of the contours of the face.
        Plast Reconstr Surg. 1978; 61: 888-891
        • Rose RM
        • Paul IL
        • Weightman B
        • Simon SR
        • Radin EL
        The role of stress-enhanced reactivity in failure of orthopaedic implants.
        J Biomed Mater Res. 1973; 4: 401-418
        • Rosenthal DI
        • Rosenberg AE
        • Schiller AL
        • Smith RJ
        Destructive arthritis due to silicone: A foreign-body reaction.
        Radiology. 1983; 149: 69-72
        • Smahel J
        Foreign material in the capsules around breast prostheses and the cellular reaction to it.
        Br J Plast Surg. 1979; 32: 35-42
        • Weightman B
        • Simon S
        • Rose R
        • Paul I
        • Radin E
        Environmental fatigue testing of silastic finger joint prostheses.
        J Biomed Mater Res. 1972; 3: 15-24
        • Worsing RA
        • Engber WD
        • Lange TA
        Reactive synovitis from particulate silastic.
        J Bone Joint Surg [Am]. 1982; 64: 581-585
        • Mayhall WST
        • Tiley FT
        • Paluska DJ
        Fracture of silastic radial-head prosthesis.
        J Bone Joint Surg [Am]. 1981; 63: 459-460
        • Sollitto RJ
        • Shonkweiler W
        Silicone shard formation: A product of implant arthroplasty.
        J Foot Surg. 1984; 23: 362-365
        • Smith RJ
        • Atkinson RE
        • Jupiter JB
        Silicone synovitis of the wrist.
        J Hand Surg. 1985; 10A: 47-61
        • Braun RM
        Stabilization of silastic implant arthroplasty at the trapezometacarpal joint.
        Clin Orthop. 1976; 121: 263-270
        • Brown JB
        • Fryer MP
        • Ohlwiler DA
        Study and use of synthetic materials, such as silicones and teflon, as subcutaneous prostheses.
        Plast Reconstr Surg. 1960; 26: 264-279
        • Gillespie TE
        • Flatt AE
        • Youm Y
        • Sprague BL
        Biomedical evaluation of metacarpophalangeal joint prosthesis designs.
        J Hand Surg. 1979; 4: 508-521
        • Goldberg AJ
        • Craig RG
        • Filisko FE
        Tear energy of elastomers for maxillofacial applications.
        J Oral Rehabil. 1980; 7: 445-551
        • Harris WH
        • Scholler AL
        • Scholler JM
        • Freiberg RA
        • Scott R
        Extensive localized bone resorption in the femur following total hip replacement.
        J Bone Joint Surg [Am]. 1976; 58: 612-618
        • Leyshon RL
        • Channon GM
        • Jenkins DHR
        • Ralis ZA
        Flexible carbon fibre in late ligamentous reconstruction for instability of the knee.
        J Bone Joint Surg [Br]. 1984; 66: 196-200
        • Coleman DL
        • King RN
        • Andrade JD
        The foreign body reaction: A chronic inflammatory response.
        J Biomed Mater Res. 1974; 8: 199-211
        • Eskeland G
        • Eskeland T
        • Hovig T
        • Teigland J
        The ultrastructure of normal digital flexor tendon sheath and of the tissue formed around silicone and polyethylene implants in man.
        J Bone Joint Surg [Br]. 1977; 59: 206-212
        • Neuman Z
        • Ben-Hur N
        • Tritsch E
        Induction of tendon sheath formation by the implantation of silicone tubes in rabbits.
        Br J Plast Surg. 1966; 19: 313-316
        • Smahel J
        • Meyer V
        Structure of capsules around silicone implants in hand surgery.
        Hand. 1983; 15: 47-52
        • Smahel J
        Histology of the capsules causing constrictive fibrosis around breast implants.
        Br J Plast Surg. 1977; 30: 324-329
        • Urbaniak JR
        • Bright DS
        • Gill LH
        • Goldner JL
        Vascularization and the gliding mechanism of free flexor-tendon grafts inserted by the silicone rod method.
        J Bone Joint Surg [Am]. 1974; 56: 473-482
        • Vistnes LM
        • Bentley JW
        • Fogarty DC
        Experimental study of tissue response to ruptured gel-filled mammary prostheses.
        Plast Reconstr Surg. 1977; 59: 31-34
        • Vistnes LM
        • Ksander GA
        • Kosek J
        Study of encapsulation of silicone rubber implants in animals: A foreignbody reaction.
        Plast Reconstr Surg. 1978; 62: 580-588
        • Rayner CRW
        The origin and nature of pseudosynovium appearing around implanted silastic rods: An experimental study.
        Hand. 1976; 8: 101-109
        • Bullough PG
        Letters to the editor (reply).
        J Bone Joint Surg [Am]. 1983; 65: 281
        • Willert HG
        • Semlitsch M
        Gschwend N Debrunner HU Total hip prosthesis. The Williams & Wilkins Co, Baltimore1976: 205-239
        • Haber J
        • Boswick JA
        • Phelps DB
        The role of soft tissue reconstruction in flexible implant arthroplasty of the metacarpophalangeal joint.
        Clin Orthop. 1979; 140: 178-183
        • Madden JW
        • DeVose G
        • Arem AJ
        A rational postoperative management program for metacarpophalangeal joint implant arthroplasty.
        J Hand Surg. 1977; 2: 358-366
        • Kleinert JM
        • Zenni Jr, EJ
        Non-union of the scaphoid: Review of literature and current treatment.
        Orthop Rev. 1984; 13: 19-35
        • Mack GR
        • Bosse MJ
        • Gelberman RH
        • Yu E
        The natural history of scaphoid nonunion.
        J Bone Joint Surg [Am]. 1984; 66: 504-509
        • le Nobel J
        • Patterson FP
        Guepar total knee prosthesis: Experience at the Vancouver General Hospital.
        J Bone Joint Surg [Br]. 1981; 63: 257-260
        • Watson HK
        • Ryu J
        • Dibella A
        An approach to Kienbock's disease: Triscaphe arthrodesis.
        J Hand Surg. 1985; 10A: 179-187
        • Pazzaglia V
        • Byers PD
        Fractured femoral shaft through an osteolytic lesion resulting from the reaction to a prosthesis.
        J Bone Joint Surg [Br]. 1984; 66: 337-339
        • Hagert CG
        Implants designed for finger joints.
        Scand J Plast Reconstr Surg. 1975; 9: 53-63
        • Engber WD
        Letters to the editor (reply).
        J Bone Joint Surg [Am]. 1983; 65: 281
        • Gordon M
        • Bullough PG
        Synovial and osseous inflammation in failed silicone-rubber prostheses.
        J Bone Joint Surg [Am]. 1982; 64: 574-580
        • Ben-Hur N
        • Ballantyne Jr, DL
        • Rees TD
        • Seidman I
        Local and systemic effects of dimethylpolysiloxane fluid in mice.
        Plast Reconstr Surg. 1967; 39: 423-426
        • Wintsch W
        • Smahel J
        • Clodius L
        Local and regional lymph node response to ruptured gel-filled mammary prostheses.
        Br J Plast Surg. 1978; 31: 349-352
        • Weeks PM
        • Vannier MW
        • Stevens WG
        • Gayou D
        • Gilula LA
        Three-dimensional imaging of the wrist.
        J Hand Surg. 1985; 10A: 32-39
        • Shapiro J
        • Laskin RS
        Giant bone cysts in rheumatoid arthritis.
        Orthop Rev. 1984; 13: 69-72
        • Wickham MG
        • Rudolph R
        • Abraham JL
        Silicon identification in prosthesis-associated fibrous capsules.
        Science. 1978; 199: 437-479