Effect of Radiocapitellar Joint Over/Under Stuffing on Elbow Joint Contact Pressure

Published:October 10, 2022DOI:https://doi.org/10.1016/j.jhsa.2021.11.006

      Purpose

      Comminuted radial head fractures are commonly treated by surgical resection or replacement with a prosthesis. A potential problem with radial head replacement is overlengthening of the radial neck (“overstuffing” of the radial head), which has been shown to affect both ulnohumeral kinematics and radiocapitellar pressures. We hypothesized that an overstuffed radial head prosthesis increases capitellar pressure and reduces coronoid pressure.

      Methods

      Seven human cadaveric elbows were prepared on a custom-designed apparatus simulating stabilizing muscle loads, and passively flexed from 0° to 90° under gravity valgus torque while joint contact pressures were measured. Each elbow was tested sequentially with different neck lengths, starting with the intact specimen followed by insertion of understuffed (−2 mm), standard-height (0 mm), and overstuffed (+2 mm) radial head prostheses in neutral forearm rotation, 40° pronation, and 40° supination positions, respectively.

      Results

      Capitellar mean contact pressures significantly increased after insertion of an overstuffed radial head prosthesis. In valgus position with neutral forearm rotation, capitellar mean contact pressure on the joint with an intact radial head averaged 227 ± 70 kPa. Insertion of understuffed, standard-height, and overstuffed radial head prostheses changed the mean contact pressures to 152 ± 76 kPa, 212 ± 68 kPa, and 491 ± 168 kPa, respectively. The overstuffed radial head group had significantly lower whole coronoid mean contact pressures (153 ± 56 kPa) compared with the intact (390 ± 138 kPa) and standard-height (376 ± 191 kPa) radial head groups.

      Conclusions

      An increase in radial prosthesis height significantly increases capitellar contact pressures and reduces coronoid contact pressures.

      Clinical relevance

      Restoration of the anatomic radial head height is critical when performing radial head arthroplasty to maintain normal joint biomechanics. Elevated capitellar contact pressures can potentially lead to pain and early degenerative changes.

      Key words

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Hand Surgery
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Mason M.L.
        Some observations on fractures of the head of the radius with a review of one hundred cases.
        Br J Surg. 1954; 42: 123-132
        • Herbertsson P.
        • Josefsson P.O.
        • Hasserius R.
        • Besjakov J.
        • Nyqvist F.
        • Karlsson M.K.
        Fractures of the radial head and neck treated with radial head excision.
        J Bone Joint Surg Am. 2004; 86: 1925-1930
        • Herbertsson P.
        • Josefsson P.O.
        • Hasserius R.
        • Karlsson C.
        • Besjakov J.
        • Karlsson M.K.
        Displaced Mason type I fractures of the radial head and neck in adults: a fifteen- to thirty-three-year follow-up study.
        J Shoulder Elbow Surg. 2005; 14: 73-77
        • Grewal R.
        • MacDermid J.C.
        • Faber K.J.
        • Drosdowech D.S.
        • King G.J.
        Comminuted radial head fractures treated with a modular metallic radial head arthroplasty. Study of outcomes.
        J Bone Joint Surg Am. 2006; 88: 2192-2200
        • Pomianowski S.
        • Morrey B.F.
        • Neale P.G.
        • Park M.J.
        • O'Driscoll S.W.
        • An K.N.
        Contribution of monoblock and bipolar radial head prostheses to valgus stability of the elbow.
        J Bone Joint Surg Am. 2001; 83: 1829-1834
        • Beingessner D.M.
        • Dunning C.E.
        • Gordon K.D.
        • Johnson J.A.
        • King G.J.
        The effect of radial head excision and arthroplasty on elbow kinematics and stability.
        J Bone Joint Surg Am. 2004; 86: 1730-1739
        • Ring D.
        Open reduction and internal fixation of fractures of the radial head.
        Hand Clin. 2004; 20 (vi): 415-427
        • Van Glabbeek F.
        • Van Riet R.P.
        • Baumfeld J.A.
        • et al.
        Detrimental effects of overstuffing or understuffing with a radial head replacement in the medial collateral-ligament deficient elbow.
        J Bone Joint Surg Am. 2004; 86: 2629-2635
        • Van Glabbeek F.
        • van Riet R.P.
        • Baumfeld J.A.
        • et al.
        The kinematic importance of radial neck length in radial head replacement.
        Med Eng Phys. 2005; 27: 336-342
        • Lanting B.A.
        • Ferreira L.M.
        • Johnson J.A.
        • King G.J.
        • Athwal G.S.
        The effect of radial head implant length on radiocapitellar articular properties and load transfer within the forearm.
        J Orthop Trauma. 2014; 28: 348-353
        • Bellato E.
        • Kim Y.
        • Fitzsimmons J.S.
        • et al.
        Coronoid reconstruction using osteochondral grafts: a biomechanical study.
        J Shoulder Elbow Surg. 2017; 26: 1794-1802
        • Drewniak E.I.
        • Crisco J.J.
        • Spenciner D.B.
        • Fleming B.C.
        Accuracy of circular contact area measurements with thin-film pressure sensors.
        J Biomech. 2007; 40: 2569-2572
        • Brimacombe J.M.
        • Wilson D.R.
        • Hodgson A.J.
        • Ho K.C.
        • Anglin C.
        Effect of calibration method on Tekscan sensor accuracy.
        J Biomech Eng. 2009; 131034503
        • Niosi C.A.
        • Wilson D.C.
        • Zhu Q.
        • Keynan O.
        • Wilson D.R.
        • Oxland T.R.
        The effect of dynamic posterior stabilization on facet joint contact forces: an in vitro investigation.
        Spine (Phila Pa 1976). 2008; 33: 19-26
        • Bellato E.
        • Fitzsimmons J.S.
        • Kim Y.
        • et al.
        Articular contact area and pressure in posteromedial rotatory instability of the elbow.
        J Bone Joint Surg Am. 2018; 100: e34
        • Hwang J.T.
        • Kim Y.
        • Bachman D.R.
        • et al.
        Axial load transmission through the elbow during forearm rotation.
        J Shoulder Elbow Surg. 2018; 27: 530-537
        • Hwang J.T.
        • Kim Y.
        • Shields M.N.
        • et al.
        Effects of axial forearm instability on force transmission across the elbow.
        J Shoulder Elbow Surg. 2019; 28: 170-177
        • Sahu D.
        • Fitzsimmons J.S.
        • Thoreson A.R.
        • An K.N.
        • O'Driscoll S.W.
        Radiocapitellar contact characteristics during prosthetic radial head subluxation.
        J Shoulder Elbow Surg. 2017; 26: 170-177
        • Sahu D.
        • Holmes D.M.
        • Fitzsimmons J.S.
        • et al.
        Influence of radial head prosthetic design on radiocapitellar joint contact mechanics.
        J Shoulder Elbow Surg. 2014; 23: 456-462
        • Bellato E.
        • Kim Y.
        • Fitzsimmons J.S.
        • et al.
        Role of the lateral collateral ligament in posteromedial rotatory instability of the elbow.
        J Shoulder Elbow Surg. 2017; 26: 1636-1643
        • Schofield J.S.
        • Evans K.R.
        • Hebert J.S.
        • Marasco P.D.
        • Carey J.P.
        The effect of biomechanical variables on force sensitive resistor error: implications for calibration and improved accuracy.
        J Biomech. 2016; 49: 786-792
        • Bachman D.R.
        • Park S.E.
        • Thaveepunsan S.
        • Fitzsimmons J.S.
        • An K.N.
        • O'Driscoll S.W.
        Joint contact changes with undersized prosthetic radial heads.
        J Orthop Trauma. 2018; 32: e440-e444
        • Fern S.E.
        • Owen J.R.
        • Ordyna N.J.
        • Wayne J.S.
        • Boardman III, N.D.
        Complex varus elbow instability: a terrible triad model.
        J Shoulder Elbow Surg. 2009; 18: 269-274
        • Bell S.N.
        • Morrey B.F.
        • Bianco A.J.
        Chronic posterior subluxation and dislocation of the radial head.
        J Bone Joint Surg Am. 1991; 73: 392-396
        • Moungondo F.
        • El Kazzi W.
        • van Riet R.
        • Feipel V.
        • Rooze M.
        • Schuind F.
        Radiocapitellar joint contacts after bipolar radial head arthroplasty.
        J Shoulder Elbow Surg. 2010; 19: 230-235
        • Hwang J.T.
        • Shields M.N.
        • Berglund L.J.
        • Hooke A.W.
        • Fitzsimmons J.S.
        • O’Driscoll S.W.
        The role of the posterior bundle of the medial collateral ligament in posteromedial rotatory instability of the elbow.
        Bone Joint J. 2018; 100-B: 1060-1065
        • An K.N.
        • Hui F.C.
        • Morrey B.F.
        • Linscheid R.L.
        • Chao E.Y.-S.
        Muscles across the elbow joint: a biomechanical analysis.
        J Biomech. 1981; 14: 659-669
        • Özkaya N.
        • Nordin M.
        Fundamentals of Biomechanics: Equilibrium, Motion, and Deformation.
        Van Nostrand Reinhold, 1991
        • Berkmortel C.J.
        • Gladwell M.S.
        • Ng J.
        • et al.
        Effect of radial neck length on joint loading.
        J Shoulder Elbow Arthroplast. 2019; 3
        • Mikíc Z.D.
        • Vukadinovíc S.M.
        Late results in fractures of the radial head treated by excision.
        Clin Orthop Relat Res. 1983; 181: 220-228
        • Ramazanian T.
        • Müller-Lebschi J.A.
        • Chuang M.Y.
        • Vaichinger A.M.
        • Fitzsimmons J.S.
        • O'Driscoll S.W.
        Effect of radiocapitellar Achilles disc arthroplasty on coronoid and capitellar contact pressures after radial head excision.
        J Shoulder Elbow Surg. 2018; 27: 1785-1791
        • Van Riet R.P.
        • Van Glabbeek F.
        • de Weerdt W.
        • Oemar J.
        • Bortier H.
        Validation of the lesser sigmoid notch of the ulna as a reference point for accurate placement of a prosthesis for the head of the radius: a cadaver study.
        J Bone Joint Surg Br. 2007; 89: 413-416
        • Nielsen K.K.
        • Olsen B.S.
        No stabilizing effect of the elbow joint capsule. A kinematic study.
        Acta Orthop Scand. 1999; 70: 6-8
        • Pollock J.W.
        • Brownhill J.
        • Ferreira L.M.
        • McDonald C.P.
        • Johnson J.A.
        • King G.J.
        Effect of the posterior bundle of the medial collateral ligament on elbow stability.
        J Hand Surg Am. 2009; 34: 116-123
        • Bell T.H.
        • Ferreira L.M.
        • McDonald C.P.
        • Johnson J.A.
        • King G.J.
        Contribution of the olecranon to elbow stability: an in vitro biomechanical study.
        J Bone Joint Surg Am. 2010; 92: 949-957