Purpose
Comminuted radial head fractures are commonly treated by surgical resection or replacement
with a prosthesis. A potential problem with radial head replacement is overlengthening
of the radial neck (“overstuffing” of the radial head), which has been shown to affect
both ulnohumeral kinematics and radiocapitellar pressures. We hypothesized that an
overstuffed radial head prosthesis increases capitellar pressure and reduces coronoid
pressure.
Methods
Seven human cadaveric elbows were prepared on a custom-designed apparatus simulating
stabilizing muscle loads, and passively flexed from 0° to 90° under gravity valgus
torque while joint contact pressures were measured. Each elbow was tested sequentially
with different neck lengths, starting with the intact specimen followed by insertion
of understuffed (−2 mm), standard-height (0 mm), and overstuffed (+2 mm) radial head
prostheses in neutral forearm rotation, 40° pronation, and 40° supination positions,
respectively.
Results
Capitellar mean contact pressures significantly increased after insertion of an overstuffed
radial head prosthesis. In valgus position with neutral forearm rotation, capitellar
mean contact pressure on the joint with an intact radial head averaged 227 ± 70 kPa.
Insertion of understuffed, standard-height, and overstuffed radial head prostheses
changed the mean contact pressures to 152 ± 76 kPa, 212 ± 68 kPa, and 491 ± 168 kPa,
respectively. The overstuffed radial head group had significantly lower whole coronoid
mean contact pressures (153 ± 56 kPa) compared with the intact (390 ± 138 kPa) and
standard-height (376 ± 191 kPa) radial head groups.
Conclusions
An increase in radial prosthesis height significantly increases capitellar contact
pressures and reduces coronoid contact pressures.
Clinical relevance
Restoration of the anatomic radial head height is critical when performing radial
head arthroplasty to maintain normal joint biomechanics. Elevated capitellar contact
pressures can potentially lead to pain and early degenerative changes.
Key words
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to Journal of Hand SurgeryAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- Some observations on fractures of the head of the radius with a review of one hundred cases.Br J Surg. 1954; 42: 123-132
- Fractures of the radial head and neck treated with radial head excision.J Bone Joint Surg Am. 2004; 86: 1925-1930
- Displaced Mason type I fractures of the radial head and neck in adults: a fifteen- to thirty-three-year follow-up study.J Shoulder Elbow Surg. 2005; 14: 73-77
- Comminuted radial head fractures treated with a modular metallic radial head arthroplasty. Study of outcomes.J Bone Joint Surg Am. 2006; 88: 2192-2200
- Contribution of monoblock and bipolar radial head prostheses to valgus stability of the elbow.J Bone Joint Surg Am. 2001; 83: 1829-1834
- The effect of radial head excision and arthroplasty on elbow kinematics and stability.J Bone Joint Surg Am. 2004; 86: 1730-1739
- Open reduction and internal fixation of fractures of the radial head.Hand Clin. 2004; 20 (vi): 415-427
- Detrimental effects of overstuffing or understuffing with a radial head replacement in the medial collateral-ligament deficient elbow.J Bone Joint Surg Am. 2004; 86: 2629-2635
- The kinematic importance of radial neck length in radial head replacement.Med Eng Phys. 2005; 27: 336-342
- The effect of radial head implant length on radiocapitellar articular properties and load transfer within the forearm.J Orthop Trauma. 2014; 28: 348-353
- Coronoid reconstruction using osteochondral grafts: a biomechanical study.J Shoulder Elbow Surg. 2017; 26: 1794-1802
- Accuracy of circular contact area measurements with thin-film pressure sensors.J Biomech. 2007; 40: 2569-2572
- Effect of calibration method on Tekscan sensor accuracy.J Biomech Eng. 2009; 131034503
- The effect of dynamic posterior stabilization on facet joint contact forces: an in vitro investigation.Spine (Phila Pa 1976). 2008; 33: 19-26
- Articular contact area and pressure in posteromedial rotatory instability of the elbow.J Bone Joint Surg Am. 2018; 100: e34
- Axial load transmission through the elbow during forearm rotation.J Shoulder Elbow Surg. 2018; 27: 530-537
- Effects of axial forearm instability on force transmission across the elbow.J Shoulder Elbow Surg. 2019; 28: 170-177
- Radiocapitellar contact characteristics during prosthetic radial head subluxation.J Shoulder Elbow Surg. 2017; 26: 170-177
- Influence of radial head prosthetic design on radiocapitellar joint contact mechanics.J Shoulder Elbow Surg. 2014; 23: 456-462
- Role of the lateral collateral ligament in posteromedial rotatory instability of the elbow.J Shoulder Elbow Surg. 2017; 26: 1636-1643
- The effect of biomechanical variables on force sensitive resistor error: implications for calibration and improved accuracy.J Biomech. 2016; 49: 786-792
- Joint contact changes with undersized prosthetic radial heads.J Orthop Trauma. 2018; 32: e440-e444
- Complex varus elbow instability: a terrible triad model.J Shoulder Elbow Surg. 2009; 18: 269-274
- Chronic posterior subluxation and dislocation of the radial head.J Bone Joint Surg Am. 1991; 73: 392-396
- Radiocapitellar joint contacts after bipolar radial head arthroplasty.J Shoulder Elbow Surg. 2010; 19: 230-235
- The role of the posterior bundle of the medial collateral ligament in posteromedial rotatory instability of the elbow.Bone Joint J. 2018; 100-B: 1060-1065
- Muscles across the elbow joint: a biomechanical analysis.J Biomech. 1981; 14: 659-669
- Fundamentals of Biomechanics: Equilibrium, Motion, and Deformation.Van Nostrand Reinhold, 1991
- Effect of radial neck length on joint loading.J Shoulder Elbow Arthroplast. 2019; 3
- Late results in fractures of the radial head treated by excision.Clin Orthop Relat Res. 1983; 181: 220-228
- Effect of radiocapitellar Achilles disc arthroplasty on coronoid and capitellar contact pressures after radial head excision.J Shoulder Elbow Surg. 2018; 27: 1785-1791
- Validation of the lesser sigmoid notch of the ulna as a reference point for accurate placement of a prosthesis for the head of the radius: a cadaver study.J Bone Joint Surg Br. 2007; 89: 413-416
- No stabilizing effect of the elbow joint capsule. A kinematic study.Acta Orthop Scand. 1999; 70: 6-8
- Effect of the posterior bundle of the medial collateral ligament on elbow stability.J Hand Surg Am. 2009; 34: 116-123
- Contribution of the olecranon to elbow stability: an in vitro biomechanical study.J Bone Joint Surg Am. 2010; 92: 949-957
Article info
Publication history
Published online: October 10, 2022
Accepted:
November 17,
2021
Received:
March 1,
2019
Publication stage
In Press Corrected ProofFootnotes
Dr O'Driscoll and the Mayo Foundation receive royalties from Acumed, LLC, Wright Medical, Inc, and Aircast, Inc. No benefits in any form have been received or will be received by the other authors related directly or indirectly to the subject of this article.
Identification
Copyright
© 2021 by the American Society for Surgery of the Hand. All rights reserved.