Comparative Accuracy of 1.5T MRI, 3T MRI, and Static Ultrasound in Diagnosis of Small Gaps in Repaired Flexor Tendons: A Cadaveric Study

Published:January 12, 2021DOI:https://doi.org/10.1016/j.jhsa.2020.10.031

      Purpose

      We hypothesized that magnetic resonance imaging (MRI) would more accurately diagnose small gaps (<6 mm) after flexor tendon repair than static ultrasound (US) and that suture artifact would negatively impair accuracy.

      Methods

      A laceration of the flexor digitorum profundus was created in 160 fresh-frozen cadaveric digits and randomized to either an intact repair (0-mm gap) or repairs using a locked 4-strand suture repair with either 4-0 Prolene, Ethibond, or and gaps of 2, 4,or 6 mm; or no suture in which 2-, 4-, or 6-mm gaps were created without a suture crossing the repair site. We performed 1.5T and 3T MRI and static US studies; gap widths were estimated by radiologists blinded to suture presence and true gap widths.

      Results

      The 1.5 and 3.0T MRI had a lower mean error than US for gap sizes 0 and 2 mm. All 3 modalities performed similarly for 4- and 6-mm gaps. Documentation of imaging artifact worsened error, and odds of seeing artifacts were 1.72 higher with MRI than with US. Suture did not worsen artifact nor impair accuracy for any of the 3 modalities. When no suture was used, all 3 modalities significantly overestimated the true gap.

      Conclusions

      MRI is most accurate for small gaps less than 4 mm. Although all modalities overestimated gap sizes in specimens with a 0-mm gap (intact tendon repair), mean overestimation (<2 mm) was not clinically relevant. Ultrasound overestimated 2-mm gaps (clinically intact repairs), whereas MRIs did not. We recommend MRI for evaluation of gaps after flexor tendon repair. The 1.5T has slightly better sensitivity and specificity for distinguishing clinically intact (gap < 3 mm) from clinically impaired (gap > 3 mm) repairs than the 3T.

      Clinical relevance

      Accurate diagnosis of intact repairs or small gaps (<3 mm) might prevent unnecessary exploration or allow modification of rehabilitation protocols. Diagnosis of clinically relevant gaps (3–6 mm) may allow for earlier revision surgery before significant tendon retraction and adhesions develop, possibly necessitating a staged reconstruction.

      Key words

      Reruptures after flexor tendon repair have a reported incidence of 4% to 6%.
      • Dy C.
      • Hernandez-Soria A.
      • Ma Y.
      • Roberts T.
      • Daluiski A.
      Complications after flexor tendon repair: a systematic review and meta-analysis.
      Stronger, multistrand repairs continue to be described, but there is no evidence that this has reduced the risk of rupture. It is possible that the actual incidence is underreported because it can be difficult to make a definitive diagnosis of rerupture. Cases often develop insidiously via progressive gapping at the repair site over the first few weeks after repair
      • Drapé J.-L.
      • Silbermann-Hoffman O.
      • Houvet P.
      • et al.
      Complications of flexor tendon repair in the hand: MR imaging assessment.
      • Zhao C.
      • Amadio P.C.
      • Momose T.
      • Couvreur P.
      • Zobitz M.E.
      • An K.N.
      Effect of synergistic wrist motion on adhesion formation after repair of partial flexor digitorum profundus tendon lacerations in a canine model in vivo.
      • Gelberman R.H.
      • Boyer M.I.
      • Brodt M.D.
      • Winters S.C.
      • Silva M.J.
      The effect of gap formation at the repair site on the strength and excursion of intrasynovial flexor tendons. An experimental study on the early stages of tendon-healing in dogs.
      and may present as triggering or stiffness, with or without contracture. Clinical diagnosis of flexor tendon reruptures has been found to be unreliable, with a reported 60% accuracy rate.
      • Matloub H.S.
      • Dzwierzynski W.W.
      • Erickson S.
      • Sanger J.R.
      • Yousif N.J.
      • Muoneke V.
      Magnetic resonance imaging scanning in the diagnosis of zone II flexor tendon rupture.
      If reexploration is not performed, the diagnosis often is unconfirmed. An accurate and precise diagnosis of gap formation after repair during the early phases of postoperative rehabilitation could have an important impact on decision-making with respect to exploration and rerepair. In addition, adjustments in the rehabilitation program (change from an active to a passive range of motion protocol) might prevent progression of a small, clinically insignificant gap (<3 mm in our opinion) to frank rupture. Furthermore, if larger gaps are not recognized early, direct rerepair may not be possible and late grafting could be required.
      • Matloub H.S.
      • Dzwierzynski W.W.
      • Erickson S.
      • Sanger J.R.
      • Yousif N.J.
      • Muoneke V.
      Magnetic resonance imaging scanning in the diagnosis of zone II flexor tendon rupture.
      Therefore, there exists a need for accurate diagnosis of repair integrity when there is a clinical concern. Currently, magnetic resonance imaging (MRI) and high-resolution ultrasound (US) are the most commonly used imaging modalities for assessment of flexor tendon repair integrity. However, there is little information about the accuracy of either of these modalities, especially for evaluating gap sizes less than 10 mm. We embarked upon the current study to determine the accuracy of MRI and to compare this modality with static, high-resolution US for diagnosing small (<6-mm) tendon gaps. Secondarily, we sought to determine how accuracy might be affected by suture artifact using a contemporary, multistrand repair technique. We hypothesized that both MRI (1.5T, 3.0T) modalities would be more accurate than high-resolution US, particularly when suture was present.

      Methods

      Our primary outcome was a continuous variable, and we wished to determine its difference in a repeated-measure data setting. A sample size of 80 was calculated for 81% power to detect an effect size of 0.36 (medium effect size) using a repeated measure design with 1 within factor and with a .05 2-sided significance level. The ulnar 4 digits (thumbs excluded) from 40 completely thawed fresh=frozen cadaveric arms (amputated above elbow) were studied (160 digits). Computer randomization was used to pair digit (index, middle, ring, or little) gap at tendon repair site (0, 2, 4, or 6 mm), and suture type used for repair (4-0 Prolene, Ethibond, or FiberWire), or no suture in which a 2-, 4-, or 6-mm gap was created but no sutures were placed across it. A Brunner incision was made in each digit from the metacarpophalangeal to the distal interphalangeal joint flexion crease. The A3 pulley was excised and a complete transverse laceration was made in the flexor digitorum profundus (FDP) tendon, midway between the distal end of the A2 and the proximal end of the A4 pulleys. The flexor digitorum sublimis tendon was left intact. A 4-strand locked cruciate repair was performed, allowing 10 mm (from lacerated ends) of suture purchase in each stump
      • Lee S.K.
      • Goldstein R.Y.
      • Zingman A.
      • Terranova C.
      • Nasser P.
      • Hausman M.R.
      The effects of core suture purchase on the biomechanical characteristics of a multistrand locking flexor tendon repair: a cadaveric study.
      (Fig. 1). This repair configuration was chosen because 4-strand repairs are becoming more popular with early motion protocols, and they have a good combination of mechanical strength and minimal bulk.
      • Dy C.J.
      • Daluiski A.
      Update on zone II flexor tendon injuries.
      ,
      • Vigler M.
      • Palti R.
      • Goldstein R.
      • Patel V.P.
      • Nasser P.
      • Lee S.K.
      Biomechanical study of cross-locked cruciate versus Strickland flexor tendon repair.
      Epitendinous repairs were not performed because of the potential for a disproportionately greater amount of suture material in the tendons with larger gaps. An electronic digital caliper accurate to 0.01 mm (Westward; Grainger International, Lake Forest, IL) was used to confirm gap width. For tendons with a 0-mm gap, the suture was locked when slight bunching of the tendon ends was visualized. Repairs were performed under 3.5× loupe magnification by the senior author (K.J.R.), a board-certified hand surgeon with 25 years’ experience. After repair, digits were fully passively flexed and extended 25 times in order to stress the repair. Gaps were remeasured, ensuring that gap widths were precisely maintained, or the repair was redone. Skin wounds were closed with a running 4-0 nylon suture. FiberWire (Arthrex, Inc., Naples, FL) suture is constructed of a multistrand, ultrahigh molecular weight polyethylene (UHMWPE) core with a braided jacket of polyester and UHMWPE. Ethibond (Ethicon Inc., 389065, Wound Closure Manual, 2007; Data on File) is a nonabsorbable, braided, suture prepared from fibers of high molecular weight, long-chain, linear polyesters. Prolene (Ethicon Inc., 389065, Wound Closure Manual, 2007; Data on File) sutures are nonabsorbable and composed of an isotactic crystalline stereoisomer of polypropylene, a synthetic linear polyolefin.
      Figure thumbnail gr1
      Figure 1Specimen with a 4-mm sutured gap using 4-0 Ethibond in a locked cruciate repair. Dark arrow depicts the knot outside the repair site.
      In specimens randomized to no suture, 2-, 4-, or 6-mm gaps were created between tendon ends, and widths were maintained by suturing the distal FDP stump to the A4 pulley and the proximal FDP stump to the A2 pulley (at least 1 cm from the gap in lacerated tendon) with a 4-0 nylon horizontal mattress suture (Fig. 2). All specimens were then evaluated within 24 hours after repair with both 1.5 and 3.0T MRI and US by a board-certified musculoskeletal radiologist (M.J.K) and ultrasonographer (N.D.), respectively, both of whom were blinded to suture presence/type, and gap width. For MRI evaluation, cadaver hands were positioned in the center of the magnet with the fingers in the extended position. Imaging was done on 1.5T GE Excite and 3T Siemens Sykra MR scanners utilizing a dedicated hand/wrist coil. The index through little fingers were scanned in the sagittal plane utilizing T1 (TR/TE; 664 ms/15 ms) and fat-suppressed PD (TR/TE; 300 ms/32 ms) sequences, with a slice thickness of 2 mm, 0.236-mm gap, nominal 6 cm × 10 cm field-of-view, and a 320–384 × 144–178 matrix. US examinations were performed in static mode using GE Logiq E9 ultrasound equipment (GE Healthcare, Milwaukee, WI). The ML6-15 and L8-18i transducers were used as found appropriate depending on the size of the finger being examined. Each radiologist also commented on whether they felt that there was any artifact (not quantified) on the images. Following imaging, surgical incisions were reopened and gaps remeasured to confirm that the original gap widths were maintained.
      Figure thumbnail gr2
      Figure 2Specimen with a nonsutured gap (red double arrow) in the flexor digitorum profundus (FDP) tendon. The tendon gap is maintained by sutures placed through the FDP and pulleys proximal and distal to the repair gap (white arrows).

      Statistical analysis

      Accuracy of the modality (defined as the difference between observed and true gap) was calculated and a paired t test used to calculate 95% confidence interval for gap measurement error. To investigate the impact of all factors (imaging modality, artifact identification, digit type, gap size, suture presence, and suture type) on accuracy, an overall mixed model with repeated measures was used. Interaction effects between imaging modality, gap size, and suture presence/type were also included in the overall model. Influence of gap size, suture, or imaging modality on subjective impression of image artifact was analyzed by logistic regression (GEE method) using artifact as the outcome, and with gap size, suture type, and imaging modality as predictors.

      Results

      Both 1.5 and 3.0T MRI had a lower mean error (difference between observed and true gap) than US for gap sizes 0 and 2 mm. All 3 modalities performed similarly for larger (4- and 6-mm) gaps, demonstrating improved accuracy. The 3.0T MRI had the lowest mean error for 0- and 4-mm gaps, and the 1.5T MRI had the lowest mean error for 2- and 6-mm gaps (Table 1). Main effect model results (least square means – model estimated from the model for each main effect) are summarized in Table 2, and interaction between imaging modality with gap size and suture type in Figures 3 and 4, respectively.
      Table 1Estimated Gap Summary Statistics by MRI Method and by Gap Size
      MRIStatisticsGap 0Gap 2Gap 4Gap 6
      1.5TMean (SD), total n0.3 (0.8), 412.0 (1.3), 423.7 (1.9), 436.0 (1.7), 26
      Median (range)0 (0.0 to 3.5)2.2 (0.0 to 5.4)3.8 (0.0 to 6.8)6.4 (0.0 to 8.8)
      Difference compared with actual gap difference (95% CI); P value
      Paired t test was used to calculate the difference and 95% CI.
      0.3 (0.06 to 0.58); <.050 (–0.38 to 0.40); .94–0.3 (–0.89 to 0.30); .320 (–0.70 to 0.64); .92
      3.0TMean (SD), total n0.3 (0.8), 422.1 (1.2), 423.7 (1.8), 445.6 (2), 28
      Median (range)0.0 (0.0 to 3.9)2.1 (0.0 to 4.4)3.9 (0.0 to 6.8)6.0 (0.0 to 9.3)
      Difference compared with actual gap difference (95% CI); P value
      Paired t test was used to calculate the difference and 95% CI.
      0.3 (0.04 to 0.56); <.050.04 (–0.35 to 0.42); .9–0.29 (–0.83 to 0.24); .3–0.39 (–1.15 to 0.38); .3
      USMean (SD), total n1.9 (2.1), 432.9 (2.7), 433.8 (2.4), 445.6 (2.6), 26
      Median (range)1.7 (0.0 to 8.0)2.0 (0.0 to 14.0)3.9 (0.0 to 9.2)6.0 (0.0 to 13.0)
      Difference compared with actual gap difference (95% CI); P value
      Paired t test was used to calculate the difference and 95% CI.
      1.9 (1.27 to 2.56); <.050.9 (0.08 to 1.72); <.05–0.2 (–0.92 to 0.53); .6–0.4 (–1.44 to 0.63); .4
      95% CI, 95% confidence interval.
      Paired t test was used to calculate the difference and 95% CI.
      Table 2Mixed Model Results Summary
      EffectLevelEstimate95% CIP Value
      MRI11.5T or 3.0T–0.09 (0.11)–0.31 to 0.12<.05
      US0.58 (0.14)0.30 to 0.85
      Artifact1Mild or severe0.52 (0.13)0.27 to 0.77<.05
      No–0.03 (0.13)–0.29 to 0.23
      DigitIndex0.35 (0.19)–0.02 to 0.72.70
      Middle0.27 (0.19)–0.10 to 0.64
      Ring0.30 (0.19)–0.07 to 0.67
      Little0.05 (0.19)–0.33 to 0.43
      Gap01.54 (0.19)1.16 to 1.92<.05
      20.42 (0.18)0.06 to 0.78
      4–0.27 (0.18)–0.63 to 0.09
      6–0.72 (0.24)–1.19 to –0.25
      Repaired typeEthibond0.08 (0.19)–0.29 to 0.46<.05
      FiberWire–0.05 (0.20)–0.43 to 0.34
      No suture1.39 (0.21)0.98 to 1.81
      Prolene–0.46 (0.19)–0.84 to –0.08
      Interaction MRI and gap sizeEstimated (Fig. 1)<.05
      Interaction MRI and repaired typeEstimated (Fig. 2).40
      95% CI, 95% confidence interval.
      Figure thumbnail gr3
      Figure 3Graph depicts the interaction between imaging modality (1.5 and 3.0 T versus US) and tendon gap size. As gap size increases, accuracy improves for all 3 modalities (P < .05).
      Figure thumbnail gr4
      Figure 4Graph demonstrates that all 3 modalities overestimate (P < .05) gap size when no suture is present. No real difference in gap estimates is seen with respect to the 3 suture types (Ethibond, FiberWire, and Prolene) studied.
      Table 3 shows sensitivity (ability to correctly diagnose a clinically impaired repair [gap > 3 mm]) and specificity (ability to diagnose a clinically intact repair [gap < 3 mm]) for all 3 modalities. US was less accurate (P < .05), than 1.5T or 3.0T MRI for all gap widths. However, gap width influenced (P < .05) accuracy for all 3 modalities. In Figure 3, for a true gap size of 0 mm (intact repair), US and both 1.5T and 3.0T MRI (combined because their results were similar) significantly (P < .05) overestimated gap width (difference > 0). Specificity for each modality was 90% (1.5T MRI), 88% (3.0T MRI), and 71% (US). Sensitivity was less for all 3 modalities: 78% (both 1.5T and 3T MRI), and 69% (US).
      Table 3Sensitivity and Specificity by Modality
      ModalityClinically Significant Gap (Gap 4 and 6 mm)Clinically Insignificant Gap (Gap 0 and 2 mm)Sensitivity (%)Specificity (%)
      1.5TPredicted significant gap (>3 mm)54878.390.4
      Predicted nonsignificant gap (≤3 mm)1575
      3.0TPredicted significant gap (>3 mm)561077.888.1
      Predicted nonsignificant gap (≤3 mm)1674
      USPredicted significant gap (>3 mm)482568.670.9
      Predicted nonsignificant gap (≤3 mm)2261
      When no suture was used, all 3 modalities significantly (p < .05) overestimated the true gap. When Ethibond, FiberWire, and Prolene suture were used (all sutured specimens combined and compared with the no suture group), gap estimates were not significantly different from 0, with US slightly overestimating and both 1.5T and 3.0T MRI underestimating the true gap (Fig. 4). Suture type did not affect accuracy.
      Neither gap size nor suture type influenced artifact being reported. The MRI method, however, did have a difference with frequency in which artifacts were reported (P < .05). Signal artifact frequencies and percentages by all 3 imaging modalities are summarized in Table 4. Among 160 observations using 1.5T MRI, 96 (60%) reported artifact compared with 92 (57.5%) for 3.0T MRI, and 73 (45.6%) for US. Therefore, when using 1.5T or 3.0T MRI to image small gaps after flexor tendon repair, the odds of seeing artifacts was 1.72 (95% confidence interval, 1.22–2.43) higher (P < .05) compared with that using US. Documentation of imaging artifact led to much larger error (P < .05).
      Table 4Artifacts Reported by Modality
      MethodOR (95% CI); P Value
      1.5T3.0TUS
      (n = 160)(n = 160)(n = 160)
      Artifact, n (%)
      No64 (40)68 (42.5)87 (54)1.7 (1.22–2.43); <.05
      Mild or severe96 (60)92 (57.5)73 (46)
      95% CI, 95% confidence interval; OR, odds ratio.

      Discussion

      Published literature regarding flexor tendon imaging has largely focused on confirmation of ruptures and localization of the proximal stump for surgical planning.
      • Lee D.H.
      • Robbin M.
      • Galliott R.
      • Graveman V.
      Ultrasound evaluation of flexor tendon lacerations.
      • Drapé J.-L.
      • Tardif-Chastenet de Gery S.
      • Silbermann-Hoffman O.
      • et al.
      Closed ruptures of the flexor digitorum tendons: MRI evaluation.
      • Budovec J.
      • Sudakoff G.
      • Dzwierzynski W.
      • Matloub H.
      • Sanger J.
      Sonographic differentiation of digital tendon rupture from adhesive scarring after primary surgical repair.
      • Bajaj S.
      • Pattamapaspong N.
      • Middleton W.
      • Teefey S.
      Ultrasound of the hand and wrist.
      With increasingly higher-resolution US probes, postprocessing software,
      • Sofka C.
      Ultrasound of the hand and wrist.
      and 3.0T MRI scanners, diagnostic accuracy may be improving, although objective data are lacking. This is particularly true with respect to small, but clinically relevant, gap formation after tendon repair. A critical gap width, which will likely lead to repair failure, is not well defined, but felt to be between 1 and 3 mm.
      • Gelberman R.H.
      • Boyer M.I.
      • Brodt M.D.
      • Winters S.C.
      • Silva M.J.
      The effect of gap formation at the repair site on the strength and excursion of intrasynovial flexor tendons. An experimental study on the early stages of tendon-healing in dogs.
      ,
      • Zhao C.
      • Amadio P.
      • Tanaka T.
      • et al.
      Effect of gap size on gliding resistance after flexor tendon repair.
      ,
      • Linnanmäki L.
      • Göransson H.
      • Havulinna J.
      • Sippola P.
      • Karjalainen T.
      • Leppänen O.V.
      Gap formation during cyclic testing of flexor tendon repair.
      Most reports evaluating the effects of gap size on tendon biomechanics have been done in animals or cadaveric specimens. US is relatively inexpensive and increasingly available, and there is growing interest for its use in musculoskeletal pathology
      • Jacobsen J.A.
      Musculoskeletal sonography and MR imaging. A role for both imaging methods.
      ; however, it does have limitations. Flexor tendon repair failure can be difficult to diagnose due to small distances between the distal and the proximal stumps, small size of the tendons, potential signal artifact from suture, and tendon adhesions that can prevent assessment of tendon gliding during dynamic US scanning. Overzealous dynamic scanning in the early weeks after repair can also be risky because of potential for tendon repair disruption with passive motion, particularly if tendon adhesions are present. In addition, motion of the digit during dynamic imaging can interfere with appropriate placement of the transducer against the skin and gel pad, potentially interfering with an ability to obtain adequate transverse and axial views needed for localization and measurement of lesions. Tendons with small, but clinically relevant gaps (>3 mm) after repair may still be able to glide in continuity during dynamic scanning, despite impending clinical failure. Adhesions may prevent tendon gliding. Complex injuries with associated fractures and joint injuries or after revascularization, as well as digits with significant joint contractures, can also be difficult to evaluate dynamically. No definitive advantage of dynamic over static US has been shown, and the latter is technically easier because flexor tendons are relatively superficial and run a straight course parallel to the skin.
      • Rubin D.A.
      • Kneeland J.B.
      • Kitay G.S.
      • Naranja Jr., R.J.
      Flexor tendon tears in the hand: use of MR imaging to diagnose degree of injury in a cadaver model.
      For this reason, we chose to compare static US evaluation to MRI. Zhang et al
      • Zhang G.-Y.
      • Zhuang H.-Y.
      • Wang L.-X.
      Value of high frequency ultrasonography in diagnosis and surgical repair of traumatic finger tendon ruptures.
      reported that US allowed localization of the distal end of the ruptured tendons and was consistent with surgical findings in all cases (95 of 95, 100%). However, in their study, mean gap distance between the proximal and distal ends of the complete ruptures was 9.2 ± 3.6 mm.
      • Zhang G.-Y.
      • Zhuang H.-Y.
      • Wang L.-X.
      Value of high frequency ultrasonography in diagnosis and surgical repair of traumatic finger tendon ruptures.
      Our findings confirm these—US is quite accurate in diagnosing larger gaps greater than 6 mm—but we wished to also determine its accuracy for diagnosing intact repairs and small gaps (<3 mm) that are clinically significant and to compare with MRI.
      MRI may be superior to US for tendon imaging in terms of soft tissue contrast and resolution, especially given the contemporary use of surface coils and reduced field-of-view imaging to enhance spatial resolution.
      • Beltran J.
      • Noto A.M.
      • Herman L.J.
      • Lubbers L.M.
      Tendons: high-field-strength surface coil MR imaging.
      MRI also allows for multiplanar imaging without manipulation of the digit. There are very few blinded studies that compare MRI with US,
      • Jacobsen J.A.
      Musculoskeletal sonography and MR imaging. A role for both imaging methods.
      especially when compared with the reference standard, exploration, and documentation of exact gap size. Many MRI studies evaluating repaired tendons are case-based, do not report estimated nor confirmed gap sizes, and present cases in which MRI was inconclusive and tendon repairs were found to be intact at the time of exploration.
      • Matloub H.S.
      • Dzwierzynski W.W.
      • Erickson S.
      • Sanger J.R.
      • Yousif N.J.
      • Muoneke V.
      Magnetic resonance imaging scanning in the diagnosis of zone II flexor tendon rupture.
      MRI has drawbacks including expense, accessibility, required use of specialized surface coils, claustrophobia, and movement-related or hardware artifacts, and it is contraindicated in patients with cardiac pacemakers, defibrillators, or prior cerebral aneurysm surgery.
      • Sofka C.
      Ultrasound of the hand and wrist.
      ,
      • Jacobsen J.A.
      Musculoskeletal sonography and MR imaging. A role for both imaging methods.
      Therefore, it would be helpful to have a clear understanding of the true diagnostic value.
      Biomechanical studies suggest that a gap greater than 3 mm after flexor tendon repair increases the risk of rerupture.
      • Gelberman R.H.
      • Boyer M.I.
      • Brodt M.D.
      • Winters S.C.
      • Silva M.J.
      The effect of gap formation at the repair site on the strength and excursion of intrasynovial flexor tendons. An experimental study on the early stages of tendon-healing in dogs.
      ,
      • Zhao C.
      • Amadio P.
      • Tanaka T.
      • et al.
      Effect of gap size on gliding resistance after flexor tendon repair.
      Therefore, the ability to diagnose intact repairs or small gaps (<3 mm) may prevent unnecessary surgical exploration. Conversely, accurate diagnosis of small, but clinically relevant gaps (3–6 mm) may allow for earlier revision surgery before significant tendon retraction and adhesions develop that might necessitate a staged reconstruction. Gap formation usually occurs in the first few weeks after tendon repair.
      • Bajaj S.
      • Pattamapaspong N.
      • Middleton W.
      • Teefey S.
      Ultrasound of the hand and wrist.
      Zhao et al
      • Zhao C.
      • Amadio P.
      • Tanaka T.
      • et al.
      Effect of gap size on gliding resistance after flexor tendon repair.
      concluded that there is a “threshold effect of 2 to 3 mm, beyond which catastrophic failure of the repair was likely” if motion continued. Therefore, accurate assessment of repair integrity in the weeks after repair, particularly if clinical assessment of motion is concerning (mismatch of active and passive motion), would be helpful during rehabilitation and may prevent unnecessary explorations, which can place neurovascular structures at risk and promote adhesions. In addition, it may help to modify rehabilitation protocols. Digital stiffness after flexor tendon repair may be due to adhesive scarring. If repair integrity can be accurately confirmed, more rigorous hand therapy may improve tendon gliding and range of motion. Similarly, therapy can be restricted (switched from active to passive motion) if a small (<3-mm) gap is identified. Popular clinical methods of assessment such as the Strickland formula are not reliable for early assessment of tendon repair integrity.
      • Matloub H.S.
      • Dzwierzynski W.W.
      • Erickson S.
      • Sanger J.R.
      • Yousif N.J.
      • Muoneke V.
      Magnetic resonance imaging scanning in the diagnosis of zone II flexor tendon rupture.
      ,
      • Corduff N.
      • Jones R.
      • Ball J.
      The role of ultrasound in the management of zone 1 flexor tendon injuries.
      We hypothesized that suture may produce echogenic foci on US or signal artifact on MRI that could be confused with a repair gap or prevent clear distinction of tendon ends. In a recent survey of members of the American Society for Surgery of the Hand, a majority reported using a 4-strand repair; with 3-0 or 4-0 core braided, nonabsorbable sutures, which encouraged us to use this configuration.
      • Gibson P.D.
      • Sobol G.L.
      • Ahmed I.H.
      Zone II flexor tendon repairs in the United States: trends in current management.
      We also used 3 different types of suture material that are popular for flexor tendon repair. Signal artifact was reported more commonly with both 1.5T and 3.0T MRI than with US, and when artifact was reported it significantly reduced accuracy of gap estimation. Interestingly, neither the presence of suture nor suture type had an effect on frequency of artifact being reported compared with gaps with no suture. Gap size also had no effect on artifact reporting. Some authors have reported poor success in diagnosing tendon ruptures in the little finger owing to the more-diminutive tendon size and have even recommended against using US to image tendons in this digit.
      • Sofka C.
      Ultrasound of the hand and wrist.
      ,
      • Ravnic D.
      • Galiano R.
      • Bodavula V.
      • Friedman D.
      • Flores R.
      Diagnosis and localisation of flexor tendon injuries by surgeon-performed ultrasound: a cadaveric study.
      Our results, however, do not show less accurate gap assessment for any of the 3 imaging modalities in the little finger, nor any significant effect related to specific digit involvement.
      Shortcomings of our study include a lack of soft tissue edema and hemorrhage in cadaver tendons, as has been shown surrounding postsurgical reruptures in previous MRI studies.
      • Corduff N.
      • Jones R.
      • Ball J.
      The role of ultrasound in the management of zone 1 flexor tendon injuries.
      Air artifact from a freshly closed incision, ice crystals, metallic debris created during tendon laceration, as well as any tendon dehydration may have adversely affected both MRI signal and US. Omission of an epitendinous suture could potentially have a negative impact on clinical applicability; however, the presence of suture was not found to decrease accuracy with any of the modalities compared with no suture. Whereas dynamic US evaluation is often helpful for distinguishing adhesions from rupture, we only used static imaging because our primary goals were to identify any adverse effect on accuracy due to gap size, suture presence and/or type, and digit size.
      In conclusion, US was less accurate than both MRI modalities. False-positive results were seen with all 3 modalities at 0-mm gaps; however, the mean error was less than 2 mm, so would most likely be interpreted as intact repairs clinically. If a 3-mm gap is considered a clinical failure, MRI can correctly diagnose this in about 80% of cases versus about 70% for US. It is important to recognize that the maximum gap size in our specimens was 6 mm. Sensitivity may, therefore, be improved for gaps larger than 6 mm, which is felt to be common with reruptures seen in clinical practice. Specificity is quite good for MRI (88%–90%) but suboptimal for US (71%). Therefore, a repair that is interpreted as having a less than 3-mm gap (clinically intact) by MRI can be relied upon in most cases, which may help surgeons make appropriate decisions with respect to treatment. Specificity of US is likely to improve as higher-resolution probes become more available, and should be a subject of future study. Suture does not impair accuracy of small-gap diagnosis with MRI nor US. Currently, we use MRI as the primary diagnostic modality for suspected flexor tendon reruptures, although gaps interpreted between 3 and 6 mm may be falsely interpreted as intact in 22% of cases. Surgical exploration may be advisable in these cases.

      References

        • Dy C.
        • Hernandez-Soria A.
        • Ma Y.
        • Roberts T.
        • Daluiski A.
        Complications after flexor tendon repair: a systematic review and meta-analysis.
        J Hand Surg Am. 2012; 37: 543-551
        • Drapé J.-L.
        • Silbermann-Hoffman O.
        • Houvet P.
        • et al.
        Complications of flexor tendon repair in the hand: MR imaging assessment.
        Radiology. 1996; 198: 219-224
        • Zhao C.
        • Amadio P.C.
        • Momose T.
        • Couvreur P.
        • Zobitz M.E.
        • An K.N.
        Effect of synergistic wrist motion on adhesion formation after repair of partial flexor digitorum profundus tendon lacerations in a canine model in vivo.
        J Bone Joint Surg Am. 2002; 84: 78-84
        • Gelberman R.H.
        • Boyer M.I.
        • Brodt M.D.
        • Winters S.C.
        • Silva M.J.
        The effect of gap formation at the repair site on the strength and excursion of intrasynovial flexor tendons. An experimental study on the early stages of tendon-healing in dogs.
        J Bone Joint Surg Am. 1999; 81: 975-982
        • Matloub H.S.
        • Dzwierzynski W.W.
        • Erickson S.
        • Sanger J.R.
        • Yousif N.J.
        • Muoneke V.
        Magnetic resonance imaging scanning in the diagnosis of zone II flexor tendon rupture.
        J Hand Surg Am. 1996; 21: 451-455
        • Lee S.K.
        • Goldstein R.Y.
        • Zingman A.
        • Terranova C.
        • Nasser P.
        • Hausman M.R.
        The effects of core suture purchase on the biomechanical characteristics of a multistrand locking flexor tendon repair: a cadaveric study.
        J Hand Surg Am. 2010; 35: 1165-1171
        • Dy C.J.
        • Daluiski A.
        Update on zone II flexor tendon injuries.
        J Am Acad Orthop Surg. 2014; 22: 791-799
        • Vigler M.
        • Palti R.
        • Goldstein R.
        • Patel V.P.
        • Nasser P.
        • Lee S.K.
        Biomechanical study of cross-locked cruciate versus Strickland flexor tendon repair.
        J Hand Surg Am. 2008; 33: 1826-1833
        • Lee D.H.
        • Robbin M.
        • Galliott R.
        • Graveman V.
        Ultrasound evaluation of flexor tendon lacerations.
        J Hand Surg Am. 2000; 25: 236-241
        • Drapé J.-L.
        • Tardif-Chastenet de Gery S.
        • Silbermann-Hoffman O.
        • et al.
        Closed ruptures of the flexor digitorum tendons: MRI evaluation.
        Skeletal Radiol. 1998; 27: 617-624
        • Budovec J.
        • Sudakoff G.
        • Dzwierzynski W.
        • Matloub H.
        • Sanger J.
        Sonographic differentiation of digital tendon rupture from adhesive scarring after primary surgical repair.
        J Hand Surg Am. 2006; 31: 524-529
        • Bajaj S.
        • Pattamapaspong N.
        • Middleton W.
        • Teefey S.
        Ultrasound of the hand and wrist.
        J Hand Surg Am. 2009; 34: 759-760
        • Sofka C.
        Ultrasound of the hand and wrist.
        Ultrasound Q. 2014; 30: 184-192
        • Zhao C.
        • Amadio P.
        • Tanaka T.
        • et al.
        Effect of gap size on gliding resistance after flexor tendon repair.
        J Bone Joint Surg Am. 2004; 86: 2482-2488
        • Linnanmäki L.
        • Göransson H.
        • Havulinna J.
        • Sippola P.
        • Karjalainen T.
        • Leppänen O.V.
        Gap formation during cyclic testing of flexor tendon repair.
        J Hand Surg Am. 2018; 43: 570.e1-570.e8
        • Jacobsen J.A.
        Musculoskeletal sonography and MR imaging. A role for both imaging methods.
        Radiol Clin North Am. 1999; 37: 713-735
        • Rubin D.A.
        • Kneeland J.B.
        • Kitay G.S.
        • Naranja Jr., R.J.
        Flexor tendon tears in the hand: use of MR imaging to diagnose degree of injury in a cadaver model.
        AJR Am J Roentgenol. 1996; 166: 615-620
        • Zhang G.-Y.
        • Zhuang H.-Y.
        • Wang L.-X.
        Value of high frequency ultrasonography in diagnosis and surgical repair of traumatic finger tendon ruptures.
        Med Princ Pract. 2012; 21: 472-475
        • Beltran J.
        • Noto A.M.
        • Herman L.J.
        • Lubbers L.M.
        Tendons: high-field-strength surface coil MR imaging.
        Radiology. 1987; 162: 735-740
        • Corduff N.
        • Jones R.
        • Ball J.
        The role of ultrasound in the management of zone 1 flexor tendon injuries.
        J Hand Surg Br. 1994; 19: 76-80
        • Gibson P.D.
        • Sobol G.L.
        • Ahmed I.H.
        Zone II flexor tendon repairs in the United States: trends in current management.
        J Hand Surg Am. 2017; 42: e99-e108
        • Ravnic D.
        • Galiano R.
        • Bodavula V.
        • Friedman D.
        • Flores R.
        Diagnosis and localisation of flexor tendon injuries by surgeon-performed ultrasound: a cadaveric study.
        J Plast Reconst Aesthet Surg. 2011; 64: 234-239