Distal Radial Fracture Management With an Intramedullary Cage and Fragment Fixation

Open AccessPublished:June 15, 2016DOI:https://doi.org/10.1016/j.jhsa.2016.05.008

      Purpose

      To examine the outcomes associated with the treatment of distal radial fractures with an expandable intramedullary cage and fragment-specific screw fixation.

      Methods

      A prospective multicenter case series of 100 patients with a fracture of the distal radius treated with the “Cage System” was undertaken. Primary patient outcomes included Disabilities of the Arm, Shoulder, and Hand, Patient-Related Wrist/Hand Evaluation questionnaires, and adverse events associated with the device. Secondary outcomes included wrist range of motion and radiographic findings.

      Results

      Follow-up was performed at intervals of 2 weeks, 4 to 6 weeks, 12 weeks, and 1 year. Ninety-one patients were available for follow-up at 2 weeks, 87 at 4 to 6 weeks, 73 at 3 months, and 61 at 12 months. The mean Disabilities of the Arm, Shoulder, and Hand score at 3 months was 21; at 12 months, it was 9. The mean total Patient-Reported Wrist/Hand Evaluation score at 3 months was 21; at 12 months, it was 11. There were 5 adverse events (5%)—2 involving radial nerve irritation and 3 involving tendon irritation from screws. Four of these 5 patients underwent surgical intervention, a neurolysis in 1 patient and removal of screws in 3. All patients were free of adverse-event symptoms at 3 months’ follow-up. Wrist range of motion improved most rapidly in the first 12 weeks following surgery and continued to improve throughout follow-up. The fracture reduction achieved at surgery was maintained throughout the healing process. There was evidence of callus formation at the fracture as early as 2 weeks after surgery with 100% of fractures healed at 3 months.

      Conclusions

      An expandable intramedullary cage with fragment-specific screw fixation provides maintenance of fracture reduction with a low complication rate.

      Type of study/level of evidence

      Therapeutic IV.

      Key words

      Fractures of the distal radius are one of the most common fractures seen in the upper extremity, especially in the elderly with a reported incidence of 195.2 per 100,000.
      • Singer B.R.
      • McLauchlan G.J.
      • Robinson C.M.
      • Christie J.
      Epidemiology of fractures in 15000 adults: the influence of age and gender.
      • Court-Brown C.M.
      • Caesar B.
      Epidemiology of adult fractures: a review.
      • Chung K.C.
      • Spilson S.V.
      The frequency and epidemiology of hand and forearm fractures in the United States.
      • Lofthus C.M.
      • Frihagen F.
      • Meyer H.E.
      • Nordsletten L.
      • Melhuus K.
      • Falch J.A.
      Epidemiology of distal forearm fractures in Oslo.
      Although approximately 60% of distal radius fractures are managed nonsurgically, an increasing trend toward surgical management has been reported as different and improved methods of internal fixation including dorsal plating, nail plate fixation, fragment specific fixation, and volar locked plating have become available.
      • Chung K.C.
      • Shauver M.J.
      • Birkmeyer J.D.
      Trends in the United States in the treatment of distal radius fractures in the elderly.
      • Koval K.J.
      • Harrast J.J.
      • Anglen J.O.
      • Weinstein J.N.
      Fractures of the distal radius. the evolution of practice over time. Where’s the evidence?.
      • Jupiter J.B.
      Fractures of the distal end of the radius.
      The introduction of volar locked plating in the early 2000s represented a major innovation in the surgical repair of distal radius fractures.
      • Orbay J.L.
      • Fernandez D.L.
      Volar fixed-angle plate fixation for unstable distal radius fractures in the elderly patient.
      Today, volar locking plates are used to treat the majority of surgically managed distal radial fractures.
      • Orbay J.L.
      • Fernandez D.L.
      Volar fixed-angle plate fixation for unstable distal radius fractures in the elderly patient.
      Most existing procedures are relatively invasive, and hardware-related complications including tendon irritation and/or rupture, hardware failure, neurological injury, loss of reduction, delayed or nonunion, and infection are relatively common to all procedures (16%–27%).
      • Arora R.
      • Lutz M.
      • Hennerbichler A.
      • Krappinger D.
      • Espen D.
      • Gabl M.
      Complications following internal fixation of unstable distal radius fracture with a palmar locking plate.
      • Jupiter J.B.
      • Marent-Huber M.
      LCP Study Group
      Operative management of distal radial fractures with 2.4-millimeter locking plates: a multicenter prospective case series.
      • Williksen J.H.
      • Husby T.
      • Hellund J.C.
      • Kvernmo H.D.
      • Rosales C.
      • Frihagen F.
      External fixation and adjuvant pins versus volar locking plate fixation in unstable distal radius fractures: a randomized, controlled study with a 5-year follow-up.
      • Knudsen R.
      • Bahadirov Z.
      • Damborg F.
      High rate of complications following volar plating of distal radius fractures.
      • Matzon J.L.
      • Kenniston J.
      • Beredjiklian P.K.
      Hardware-related complications after dorsal plating for displaced distal radius fractures.
      Therefore, the need still exists for a surgical solution that minimizes soft tissue trauma and hardware-related complications, addresses a broad range of fracture patterns, and promotes rapid return to function. The objective of this study was to assess the functional and radiographic outcomes for a series of patients treated with an expandable metallic implant that is deployed into the medullary canal of the distal radius providing a scaffold to which bone fragments can be stabilized using fragment specific screws.

      Materials and Methods

      A multicenter nonsequential study was undertaken at multiple health centers in Germany after the study protocol was approved by the Ethics Committee in each participating center. The protocol conformed to the ethical guidelines of the 1975 Declaration of Helsinki. Required exclusion criteria for this study, in accordance with Conformité Européene marked product labeling, included mental illness, alcoholism, foreign body sensitivity, AO class C3 fractures, and open growth plates.
      Participating centers were provided financial compensation by the implant manufacturer for completion of required study follow-up only. No payments were made to physicians or patients.
      Follow-up comprising completion of the Disabilities of the Arm, Shoulder, and Hand (DASH) and Patient-Related Wrist/Hand Evaluation (PRWHE) forms, measurement of range of motion, and assessment for adverse events was performed at 2 weeks, 4 to 6 weeks, 12 weeks, and 1 year by the principal investigators (M.S., M.J., and W.S.), subinvestigators, and/or site coordinators. Of the 100 patients initially admitted to the study, 91 patients were available for follow-up at 2 weeks, 87 at 4 to 6 weeks, 73 at 12 weeks, and 61 at 1 year.
      Patients were lost to follow-up for a number of reasons including voluntary withdrawal by 26 individuals because they were satisfied with the results from their surgery and they, therefore, requested to not have to return for additional follow-up. Seven patients were lost from 1 site because the principal investigator changed practice location. Six patients, at multiple sites, were excluded because they missed follow-up appointments.
      X-rays were obtained before surgery, immediately after surgery, and at intervals up to 1 year. Follow-up x-rays were available on 63 patients at 12 weeks and 48 patients at 1 year. X-rays were taken in compliance with each site’s standard of care, which meant, in many instances, x-rays were not allowed after the fracture was judged to be healed out of concern for unnecessary radiation exposure to the patient. X-rays, when available, were evaluated in final follow-up by a board certified orthopedic hand surgeon (A.P.) in regards to radial inclination, volar tilt, ulnar variance, coronal shift, and evidence of fracture healing. Fracture union was judged to be present if osseous bridging across the fracture site was seen in 2 of 3 views as recommended by Rosental et al.
      • Rozental T.D.
      • Blazar P.E.
      • Franko O.I.
      • Chacko A.T.
      • Earp B.E.
      • Day C.S.
      Functional outcomes for unstable distal radial fractures treated with open reduction and internal fixation or closed reduction and percutaneous fixation. A prospective randomized trial.

      Surgical technique

      All surgical procedures were performed through a standard radial approach. After a reduction of the fracture was obtained by the surgeon, using either closed or open methods and held in place via longitudinal finger trap traction or axially placed K-wires, the technique and instrumentation unique to the Cage Distal Radius System (Conventus Orthopedics, Maple Grove, MN) was utilized. The technique involves entering the radial aspect of the radius using drill bits through a 2- to 3-cm radial incision that is located approximately 3 cm proximal to the articular surface. This is followed by preparation of the site for the implant with the system’s reaming device in the subcortical region of the distal radius. The implant is then deployed into the prepared space of the distal radius just proximal to the fracture. The device is locked in its expanded position and fixed to the radial border of the radius with a small side plate and screws. The final step is the introduction of fragment-specific, cannulated screws through the device, across the fracture and into the far cortex.
      Postoperative immobilization varied according to the surgeon’s preference and the site’s standard of care for distal radial fractures. Fifty-five patients underwent no immobilization, 40 utilized a commercial orthosis for 2 weeks, and in 5 patients, a cast was used for 2 weeks. No patients received formal therapy but were instructed to begin range of motion exercises as tolerated.

      Results

      Eighty-nine percent of patients were female with a mean age of 68 years. Ninety-three percent were right-handed, although the right and left wrists were fractured nearly equally. The AO classification of fractures seen included A2, A3, B1, B2, B3, C1, and C2 (Table 1).
      Table 1Demographic Data
      CharacteristicSubjects
      Age (y)n = 100
       Mean68 ± 12
       Range33–94
      Gender
       Female89% (89 of 100)
      Mechanism of injuryn = 100
       Fall96% (96 of 100)
       Sports2% (2 of 100)
       Other (MVA)2% (2 of 100)
       AO classification
      A227% (27 of 101)
      A337% (38 of 101)
      B16% (6 of 101)
      B25% (5 of 101)
      B31% (1 of 101)
      C119% (19 of 101)
      C25% (5 of 101)
      Fractured hand
       Right51% (51 of 100)
       Left49% (49 of 100)
      Dominant hand
       Right93% (93 of 100)
       Left6% (6 of 100)
      MVA, motor vehicle accident.

      Clinical outcomes

      The DASH questionnaire was completed by all patients at the time of injury with a mean score of 77. At 2 weeks, the DASH had improved to a mean of 57, at 4 to 6 weeks to a mean of 37, at 3 months to a mean of 21, and at 1 year to a mean of 9 (Fig. 1). The PRWHE questionnaire showed similar improvement in results with a decrease in a mean score of 74 at the time of the fracture to a mean of 11 at 1 year follow-up (Fig. 2).
      Figure thumbnail gr1
      Figure 1The DASH score is seen to improve over 1 year with the greatest improvement seen between the 2-week and the 4- to 6-week period.
      Figure thumbnail gr2
      Figure 2PRWHE score is seen to improve gradually over 1 year.

      Adverse events

      There were 2 patients reporting radial nerve complaints after surgery. One of these patients was treated with a neurolysis at 8 weeks. Both reported complete resolution of symptoms at the 12-week follow-up.
      Three patients developed pain associated with prominent screws during the healing period. All 3 patients reported complete resolution of symptoms following removal of 1 or more of the prominent screws after fracture healing. None of the devices required removal.

      Range of motion

      Wrist flexion and extension, radial and ulnar deviation, and pronation and supination were measured with motion expressed in terms of a percentage of the uninjured extremity with the exception of the 1 patient with bilateral wrist fractures. Figure 3 shows that, on average, flexion and extension measured about 50% of normal at 2 weeks and had improved to about 75% of normal by 12 weeks. Radial and ulnar deviation measured close to 60% at 2 weeks and had improved to 78% by 12 weeks. Pronation and supination improved to 88% of normal at 2 weeks and 95% at 12 weeks. Range of motion continued to improve up to 1 year, but the majority of improvement was seen in the first 12 weeks.
      Figure thumbnail gr3
      Figure 3A Range of motion of the injured wrist compared with the uninjured wrist improved with time in flexion/extension. B Radial/ulnar deviation. C Pronation/supination.

      Radiographic evaluation

      X-rays taken before surgery, immediately after surgery, and at intervals up to 1 year were closely evaluated. Table 2 illustrates that most fractures were displaced with a loss of radial inclination and normal volar tilt before surgery. There was also often radial shortening as measured by increased ulnar variance.
      • Graham T.J.
      Surgical Correction of malunited fractures of the distal radius.
      Coronal shift of the fracture was seen in some instances. These deformities were all corrected back to normal or near-normal anatomical parameters at the time of surgery with some surgeons choosing, when they could not get an anatomical reduction with closed methods, to accept some radial shortening and lack of complete restoration of normal volar tilt because of the patient’s age or compromised medical condition. The reduction obtained and accepted at the time of surgery was maintained throughout fracture healing.
      Table 2Radiographic Data
      StatisticPreoperativePostoperative2 Wk4-6 Wk12 Wk12 Mo
      Radial Inclination (°)
      Mean16.519.919.619.619.719.7
      Range2–2215–2215–2214–2214–2214–22
      Volar Tilt (°)
      Mean
      − refers to degrees less than neutral and + refers to degrees more than neutral.
      +13.1−1.5−1.8−1.3−1.2−1.4
      Range
      − refers to degrees less than neutral and + refers to degrees more than neutral.
      −5 to +38−5 to +5−5 to +6−5 to +6−5 to +6−5 to +6
      Ulnar Variance (mm)
      Mean
      − refers to degrees less than neutral and + refers to degrees more than neutral.
      +1.5+0.1+0.5+0.5+0.4+0.5
      Range
      − refers to degrees less than neutral and + refers to degrees more than neutral.
      −2 to +5−2 to +4−2 to +4−2 to +4−2 to +4−2 to +4
      Evidence of Fracture Healing
      BaselinePostoperative2 Wk4-6 Wk12 Wk
      NANA7.5% (3 of 40)87.5% (35 of 40)100% (40 of 40)
      − refers to degrees less than neutral and + refers to degrees more than neutral.
      Early callus formation was seen at the fracture site in 7.5% of cases at 2 weeks and in 87.5% of cases at 4 to 6 weeks. All fractures were healed at 12 weeks (Fig. 4).
      Figure thumbnail gr4
      Figure 4A Two views of a dorsally comminuted, displaced, and angulated extra-articular distal radial fracture. B The expandable cage and 2 fragment fixation screws in place at the time of surgery. The arrows show the fracture line. C Early healing radially 4 to 6 weeks after surgery. The arrows show the fracture line with presence of callus formation. D Healing of the fracture (arrows) at 3 months after surgery.

      Discussion

      Over the past decade, the operative management of distal radius fractures has moved from dorsal plating and fragment-specific fixation to the widespread use of volar plates for the reduction and treatment of simple and complex fractures. Volar plating offers several advantages over other techniques, including visualization of the volar fracture fragments and ease of plate application; however, complications such as tendon rupture, transient nerve injuries, and postoperative stiffness associated with plate placement, screw placement, and pronator quadratus stripping may occur.
      • Williksen J.H.
      • Husby T.
      • Hellund J.C.
      • Kvernmo H.D.
      • Rosales C.
      • Frihagen F.
      External fixation and adjuvant pins versus volar locking plate fixation in unstable distal radius fractures: a randomized, controlled study with a 5-year follow-up.
      • Knudsen R.
      • Bahadirov Z.
      • Damborg F.
      High rate of complications following volar plating of distal radius fractures.
      • Matzon J.L.
      • Kenniston J.
      • Beredjiklian P.K.
      Hardware-related complications after dorsal plating for displaced distal radius fractures.
      • Roh Y.H.
      • Lee B.K.
      • Baek J.R.
      • Noh J.H.
      • Gong H.S.
      • Baek G.H.
      A randomized comparison of volar plate and external fixation for intra-articular distal radius fractures.
      • Gunther S.B.
      • Lynch T.L.
      Rigid internal fixation of displaced distal radius fractures.
      Intramedullary fixation of distal radial fractures, through a minimal approach utilizing a nail, has been used with successful outcomes. However, this technique does not allow for screw fixation of separate fracture fragments to the nail construct and is best reserved for simple A-type fractures.
      • Gunther S.B.
      • Lynch T.L.
      Rigid internal fixation of displaced distal radius fractures.
      • Gradl G.
      • Mielsch N.
      • Wendt M.
      • et al.
      Intramedullary nail versus volar plate fixation of extra-articular distal radius fractures. Two year results of a prospective randomized trial.
      • Tan V.
      • Bratchenko W.
      • Nourbakhsh A.
      • Capo J.
      Comparative analysis of intramedullary nail fixation versus casting for treatment of distal radius fractures.
      • Safi A.
      • Hart R.
      • Te˘kne˘džjan B.
      • Kozák
      Treatment of extra-articular and simple articular distal radial fractures with intramedullary nail versus volar locking plate.
      • Schønnemann J.O.
      • Hansen T.B.
      • Søballe K.
      Randomised study of non-bridging external fixation compared with intramedullary fixation of unstable distal radial fractures.
      The intramedullary cage that has been shown to be biomechanically equivalent to volar plates was designed to be inserted in a collapsed state and then expanded to fill a spot created with a cavity preparation device in the subcortical region of the distal radius just proximal to the distal radial fracture.
      • van Kampen R.J.
      • Thoreson A.R.
      • Knutson N.J.
      • Hale J.E.
      • Moran S.L.
      Comparison of a new intramedullary scaffold to volar plating for treatment of distal radius fractures.
      By inserting the device in a collapsed state, the procedure may be less invasive than traditional open plating. The device itself is not intended to be used to reduce a distal radial fracture but to stabilize the fracture once a reduction has been obtained through conventional methods. In this study, the device was not used for treating highly comminuted fractures such as AO class C3.
      The advantages of the technique presented here include a small incision, minimal periosteal stripping, intramedullary placement of the fixation construct, and the ability for the surgeon to choose the number and direction of screws used to hold fracture fragments to the construct.
      The mean DASH and total PRWHE scores for all subjects for whom 12-week follow-up was obtained (n = 73) were very similar, with both being reported at 21. For those for whom there was follow-up of 1 year (n = 47), mean DASH was 9 and total PRWHE was 11. For both DASH and PRWHE, the biggest improvement was seen in the first 12 weeks following treatment. These findings compare favorably with reports in the literature and are similar to those seen with other devices for distal radial fracture fixation.
      • Schønnemann J.O.
      • Hansen T.B.
      • Søballe K.
      Randomised study of non-bridging external fixation compared with intramedullary fixation of unstable distal radial fractures.
      • Capo J.T.
      • Hashem J.
      • Orillaza N.S.
      • Tan V.
      • Warburton M.
      • Bonilla L.
      Treatment of extra-articular distal radial malunions with an intramedullary implant.
      • Egol K.
      • Walsh M.
      • Tejwani N.
      • McLaurin T.
      • Wynn C.
      • Paksima N.
      Bridging external fixation and supplementary Kirschner-wire fixation versus volar locked plating for unstable fractures of the distal radius: a randomised, prospective trial.
      • Ilyas A.M.
      • Thoder J.J.
      Intramedullary fixation of displaced distal radius fractures: a preliminary report.
      • Marcheix P.S.
      • Dotzis A.
      • Benkö P.E.
      • Siegler J.
      • Arnaud J.P.
      • Charissoux J.L.
      Extension fractures of the distal radius in patients older than 50: a prospective randomized study comparing fixation using mixed pins or a palmar fixed-angle plate.
      • McFadyen I.
      • Field J.
      • McCann P.
      • Ward J.
      • Nicol S.
      • Curwen C.
      Should unstable extra-articular distal radial fractures be treated with fixed-angle volar-locked plates or percutaneous Kirschner wires? A prospective randomised controlled trial.
      • van Manen C.J.
      • Dekker M.L.
      • van Eerten P.V.
      • Rhemrev S.J.
      • van Olden G.D.
      • van der Elst M.
      Bio-resorbable versus metal implants in wrist fractures: a randomised trial.
      • Wei D.H.
      • Raizman N.M.
      • Bottino C.J.
      • Jobin C.M.
      • Strauch R.J.
      • Rosenwasser M.P.
      Unstable distal radial fractures treated with external fixation, a radial column plate, or a volar plate. A prospective randomized trial.
      • Wilcke M.K.
      • Abbaszadegan H.
      • Adolphson P.Y.
      Wrist function recovers more rapidly after volar locked plating than after external fixation but the outcomes are similar after 1 year.
      • Hunsaker F.G.
      • Cioffi D.A.
      • Amadio P.C.
      • Wright J.G.
      • Caughlin B.
      The American Academy of Orthopaedic Surgeons Outcome Instruments. Normative values from the general population.
      • Lauder A.
      • Agnew S.
      • Bakri K.
      • Allan C.H.
      • Hanel D.P.
      • Huang J.I.
      Functional outcomes following bridge plate fixation for distal radius fractures.
      • Costa M.L.
      • Achten J.
      • Parsons N.R.
      • et al.
      Volar locking-plate and Kirschner-wire fixation did not differ in terms of functional outcomes after dorsally displaced distal radial fracture.
      A total of 5 subjects (5 of 100; 5%) experienced an adverse event related to the device and/or procedure. Two subjects experienced radial nerve symptoms and 3 subjects reported tendon irritation associated with 1 or more of the fragment-specific screws. The symptoms resolved after removal of the prominent screws in all cases. These adverse events are also seen with all other forms of internal fixation for distal radial fractures, although the incidence of 5% for adverse events associated with this device is lower than that reported with other forms of internal fixation.
      • Arora R.
      • Lutz M.
      • Hennerbichler A.
      • Krappinger D.
      • Espen D.
      • Gabl M.
      Complications following internal fixation of unstable distal radius fracture with a palmar locking plate.
      • Jupiter J.B.
      • Marent-Huber M.
      LCP Study Group
      Operative management of distal radial fractures with 2.4-millimeter locking plates: a multicenter prospective case series.
      • Williksen J.H.
      • Husby T.
      • Hellund J.C.
      • Kvernmo H.D.
      • Rosales C.
      • Frihagen F.
      External fixation and adjuvant pins versus volar locking plate fixation in unstable distal radius fractures: a randomized, controlled study with a 5-year follow-up.
      • Knudsen R.
      • Bahadirov Z.
      • Damborg F.
      High rate of complications following volar plating of distal radius fractures.
      • Matzon J.L.
      • Kenniston J.
      • Beredjiklian P.K.
      Hardware-related complications after dorsal plating for displaced distal radius fractures.
      To lessen the occurrence of radial nerve problems, we now recommend that the self-retaining retractor used in the radial incision be removed as soon as possible during the procedure. To prevent problems associated with prominent screws, we now recommend that a cut-down incision be done in conjunction with placement of each screw and that the head of each screw can be buried in the bone.
      In this series, there were no reported occurrences of infections, device failure necessitating device removal, or carpal tunnel syndrome. Such complications have been reported with other forms of internal fixation, ranging from 14% to 22%.
      • Arora R.
      • Lutz M.
      • Hennerbichler A.
      • Krappinger D.
      • Espen D.
      • Gabl M.
      Complications following internal fixation of unstable distal radius fracture with a palmar locking plate.
      • Jupiter J.B.
      • Marent-Huber M.
      LCP Study Group
      Operative management of distal radial fractures with 2.4-millimeter locking plates: a multicenter prospective case series.
      • Williksen J.H.
      • Husby T.
      • Hellund J.C.
      • Kvernmo H.D.
      • Rosales C.
      • Frihagen F.
      External fixation and adjuvant pins versus volar locking plate fixation in unstable distal radius fractures: a randomized, controlled study with a 5-year follow-up.
      • Knudsen R.
      • Bahadirov Z.
      • Damborg F.
      High rate of complications following volar plating of distal radius fractures.
      • Matzon J.L.
      • Kenniston J.
      • Beredjiklian P.K.
      Hardware-related complications after dorsal plating for displaced distal radius fractures.
      • Gradl G.
      • Mielsch N.
      • Wendt M.
      • et al.
      Intramedullary nail versus volar plate fixation of extra-articular distal radius fractures. Two year results of a prospective randomized trial.
      • McKay S.D.
      • MacDermid J.C.
      • Roth J.H.
      • Richards R.S.
      Assessment of complications of distal radius fractures and development of a complication checklist.
      • Davis D.I.
      • Baratz M.
      soft tissue complications of distal radius fractures.
      Although there were no device removals in this study, there have been cases in which the implant has been removed. In Germany, health care practitioners routinely remove metallic and other implants if they are of no additional benefit to the patient. The intramedullary device and screws have been removed up to 444 days after insertion as requested by the patient. In each case, the device was able to be removed without causing additional damage to surrounding tissues.
      Radiographic follow-up in this study revealed that the intramedullary device and fragment-specific screw fixation resulted in a stable fracture construct as evidenced by minimal or no change in position of the postoperative reduction throughout healing (Table 2). Full healing of the fracture was seen in 100% of cases at 12 weeks as well as early signs of healing in 87% of cases at 4 to 6 weeks. Because cavity preparation for the implant essentially creates a bone slurry from the intramedullary bone while removing no bone, it is possible that an osteogenic milieu is created that contributes to rapid fracture healing. Further study of this is required.
      The study shows that an expandable intramedullary implant with fragment-specific screw fixation has been associated with satisfactory results and a low complication rate when used for treating distal radius fractures, with the exception of AO class C3 fractures.

      References

        • Singer B.R.
        • McLauchlan G.J.
        • Robinson C.M.
        • Christie J.
        Epidemiology of fractures in 15000 adults: the influence of age and gender.
        J Bone Joint Surg Br. 1998; 80: 243-248
        • Court-Brown C.M.
        • Caesar B.
        Epidemiology of adult fractures: a review.
        Injury. 2006; 37: 691-697
        • Chung K.C.
        • Spilson S.V.
        The frequency and epidemiology of hand and forearm fractures in the United States.
        J Hand Surg Am. 2001; 26: 908-915
        • Lofthus C.M.
        • Frihagen F.
        • Meyer H.E.
        • Nordsletten L.
        • Melhuus K.
        • Falch J.A.
        Epidemiology of distal forearm fractures in Oslo.
        Norway. Osteoporos Int. 2008; 19: 781-786
        • Chung K.C.
        • Shauver M.J.
        • Birkmeyer J.D.
        Trends in the United States in the treatment of distal radius fractures in the elderly.
        J Bone Joint Surg Am. 2009; 91: 1868-1873
        • Koval K.J.
        • Harrast J.J.
        • Anglen J.O.
        • Weinstein J.N.
        Fractures of the distal radius. the evolution of practice over time. Where’s the evidence?.
        J Bone Joint Surg Am. 2008; 90: 1855-1861
        • Jupiter J.B.
        Fractures of the distal end of the radius.
        J Bone Joint Surg Am. 1991; 73: 461-469
        • Orbay J.L.
        • Fernandez D.L.
        Volar fixed-angle plate fixation for unstable distal radius fractures in the elderly patient.
        J Hand Surg Am. 2004; 29: 96-102
        • Arora R.
        • Lutz M.
        • Hennerbichler A.
        • Krappinger D.
        • Espen D.
        • Gabl M.
        Complications following internal fixation of unstable distal radius fracture with a palmar locking plate.
        J Orthop Trauma. 2007; 21: 316-322
        • Jupiter J.B.
        • Marent-Huber M.
        • LCP Study Group
        Operative management of distal radial fractures with 2.4-millimeter locking plates: a multicenter prospective case series.
        J Bone Joint Surg Am. 2009; 91: 55-65
        • Williksen J.H.
        • Husby T.
        • Hellund J.C.
        • Kvernmo H.D.
        • Rosales C.
        • Frihagen F.
        External fixation and adjuvant pins versus volar locking plate fixation in unstable distal radius fractures: a randomized, controlled study with a 5-year follow-up.
        J Hand Surg Am. 2015; 40: 1333-1340
        • Knudsen R.
        • Bahadirov Z.
        • Damborg F.
        High rate of complications following volar plating of distal radius fractures.
        Dan Med J. 2014; 61: A4906
        • Matzon J.L.
        • Kenniston J.
        • Beredjiklian P.K.
        Hardware-related complications after dorsal plating for displaced distal radius fractures.
        Orthopedics. 2014; 37: e978-e982
        • Rozental T.D.
        • Blazar P.E.
        • Franko O.I.
        • Chacko A.T.
        • Earp B.E.
        • Day C.S.
        Functional outcomes for unstable distal radial fractures treated with open reduction and internal fixation or closed reduction and percutaneous fixation. A prospective randomized trial.
        J Bone Joint Surg Am. 2009; 91: 1837-1846
        • Graham T.J.
        Surgical Correction of malunited fractures of the distal radius.
        J Am Acad Orthop Surg. 1997; 5: 270-281
        • Roh Y.H.
        • Lee B.K.
        • Baek J.R.
        • Noh J.H.
        • Gong H.S.
        • Baek G.H.
        A randomized comparison of volar plate and external fixation for intra-articular distal radius fractures.
        J Hand Surg Am. 2015; 40: 34-41
        • Gunther S.B.
        • Lynch T.L.
        Rigid internal fixation of displaced distal radius fractures.
        Orthopedics. 2014; 37: e34-e38
        • Gradl G.
        • Mielsch N.
        • Wendt M.
        • et al.
        Intramedullary nail versus volar plate fixation of extra-articular distal radius fractures. Two year results of a prospective randomized trial.
        Injury. 2014; 45: S3-S8
        • Tan V.
        • Bratchenko W.
        • Nourbakhsh A.
        • Capo J.
        Comparative analysis of intramedullary nail fixation versus casting for treatment of distal radius fractures.
        J Hand Surg Am. 2012; 37: 460-468
        • Safi A.
        • Hart R.
        • Te˘kne˘džjan B.
        • Kozák
        Treatment of extra-articular and simple articular distal radial fractures with intramedullary nail versus volar locking plate.
        J Hand Surg Eur Vol. 2013; 38: 774-779
        • Schønnemann J.O.
        • Hansen T.B.
        • Søballe K.
        Randomised study of non-bridging external fixation compared with intramedullary fixation of unstable distal radial fractures.
        J Plast Hand Surg. 2011; 45: 232-237
        • van Kampen R.J.
        • Thoreson A.R.
        • Knutson N.J.
        • Hale J.E.
        • Moran S.L.
        Comparison of a new intramedullary scaffold to volar plating for treatment of distal radius fractures.
        J Orthop Trauma. 2013; 27: 535-541
        • Capo J.T.
        • Hashem J.
        • Orillaza N.S.
        • Tan V.
        • Warburton M.
        • Bonilla L.
        Treatment of extra-articular distal radial malunions with an intramedullary implant.
        J Hand Surg Am. 2010; 35: 892-899
        • Egol K.
        • Walsh M.
        • Tejwani N.
        • McLaurin T.
        • Wynn C.
        • Paksima N.
        Bridging external fixation and supplementary Kirschner-wire fixation versus volar locked plating for unstable fractures of the distal radius: a randomised, prospective trial.
        J Bone Joint Surg Br. 2008; 90: 1214-1221
        • Ilyas A.M.
        • Thoder J.J.
        Intramedullary fixation of displaced distal radius fractures: a preliminary report.
        J Hand Surg Am. 2008; 33: 1706-1715
        • Marcheix P.S.
        • Dotzis A.
        • Benkö P.E.
        • Siegler J.
        • Arnaud J.P.
        • Charissoux J.L.
        Extension fractures of the distal radius in patients older than 50: a prospective randomized study comparing fixation using mixed pins or a palmar fixed-angle plate.
        J Hand Surg Eur Vol. 2010; 35: 646-651
        • McFadyen I.
        • Field J.
        • McCann P.
        • Ward J.
        • Nicol S.
        • Curwen C.
        Should unstable extra-articular distal radial fractures be treated with fixed-angle volar-locked plates or percutaneous Kirschner wires? A prospective randomised controlled trial.
        Injury. 2011; 42: 162-166
        • van Manen C.J.
        • Dekker M.L.
        • van Eerten P.V.
        • Rhemrev S.J.
        • van Olden G.D.
        • van der Elst M.
        Bio-resorbable versus metal implants in wrist fractures: a randomised trial.
        Arch Orthop Trauma Surg. 2008; 128: 1413-1417
        • Wei D.H.
        • Raizman N.M.
        • Bottino C.J.
        • Jobin C.M.
        • Strauch R.J.
        • Rosenwasser M.P.
        Unstable distal radial fractures treated with external fixation, a radial column plate, or a volar plate. A prospective randomized trial.
        J Bone Joint Surg Am. 2009; 91: 1568-1577
        • Wilcke M.K.
        • Abbaszadegan H.
        • Adolphson P.Y.
        Wrist function recovers more rapidly after volar locked plating than after external fixation but the outcomes are similar after 1 year.
        Acta Orthop. 2011; 82: 76-81
        • Hunsaker F.G.
        • Cioffi D.A.
        • Amadio P.C.
        • Wright J.G.
        • Caughlin B.
        The American Academy of Orthopaedic Surgeons Outcome Instruments. Normative values from the general population.
        J Bone Joint Surg Am. 2002; 84: 208-215
        • Lauder A.
        • Agnew S.
        • Bakri K.
        • Allan C.H.
        • Hanel D.P.
        • Huang J.I.
        Functional outcomes following bridge plate fixation for distal radius fractures.
        J Hand Surg Am. 2015; 40: 1554-1562
        • Costa M.L.
        • Achten J.
        • Parsons N.R.
        • et al.
        Volar locking-plate and Kirschner-wire fixation did not differ in terms of functional outcomes after dorsally displaced distal radial fracture.
        J Bone Joint Surg Am. 2015; 97A: 859
        • McKay S.D.
        • MacDermid J.C.
        • Roth J.H.
        • Richards R.S.
        Assessment of complications of distal radius fractures and development of a complication checklist.
        J Hand Surg Am. 2001; 26: 916-922
        • Davis D.I.
        • Baratz M.
        soft tissue complications of distal radius fractures.
        Hand Clin. 2010; 26: 229-235

      Linked Article

      • Erratum
        Journal of Hand SurgeryVol. 41Issue 10
        • Preview
          In the article by Strassmair MK, Jonas M, Schäfer W, and Palmer A in the August 2016 issue of the Journal (“Distal Radial Fracture Management With an Intramedullary Cage and Fragment Fixation”, Vol. 41, No. 8, p. 833-840), the conflict of interest statement was incorrect. Dr Palmer is a paid consultant for Conventus Orthopaedics. The publisher regrets this error.
        • Full-Text
        • PDF