Purpose

The use of locking plates increases the primary load to failure, thereby reducing the rate of implant-related failure. The good clinical and biomechanical results of locking plates in long bones might be applicable to treatment of metacarpal fractures. The purpose of this study was to determine strength and stiffness of locking plates in a metacarpal fracture model with mono- and bicortical screw fixation in comparison to non-locking plate mono- and bicortical screw fixation, with both types of plates placed at the dorsal side of the bone.

Methods

Fresh second metacarpals from domestic pigs (n=40) were randomized in 4 equal groups. Short, oblique, mid-shaft fractures were generated, using a standardized 3-point bending method. Fractures were plated with non-locking, titanium, 1-mm-thick monocortical (group 1, n=10) or bicortical (group 2, n =10) plates (Leibinger-Stryker; Stryker Corp, Freiburg, Germany). Newly designed locking titanium plates with the same width and thickness (Leibinger-Stryker) were used in the same manner for groups 3 (monocortical) and 4 (bicortical). The metacarpals were then tested to load to failure in a cantilever bending mode.

Results

Bicortical, non-locking fixation (group 2, 359 ± 90 N) had a higher load to failure than monocortical non-locking fixation (group 1, 250 ± 56 N) in testing the maximum load to failure (p < .01). There was no significant difference in stiffness between group 1 (46 ± 12 N/mm) and group 2 (56 ± 21 N/mm). The difference in maximum load to failure between monocortical (group 3, 440 ± 85N) and bicortical (group 4, 378 ± 116 N) locking plate stabilization was not significant. Also, there was no significant difference in stiffness between monocortical (group 3, 83 ± 35 N/mm) and bicortical locking plates (group 4, 70 ± 31 N/mm). Comparing non-locking (group 1) and locking plates in a monocortical fixation technique (group 3) demonstrated significant differences in maximum load to failure (group 1, 250 ± 56 N; group 3, 440 ± 85 N) and stiffness (group 1, 46 ± 12 N/mm; group 3, 83 ± 35 N/mm). The stability of monocortical locking plates was stronger, although not statistically significant, than the non-locking bicortical plates (load to failure, 440 ± 85 N vs 359 ± 90 N; stiffness, 83 ± 35 N/mm vs 56 ± 21 N/mm).

Conclusions

The new generation of locking plates can be used to achieve a higher stability for fixation of metacarpal fractures. Monocortical, stable fixation can minimize flexor tendon interference and probably reduce bone and soft tissue trauma.

Key words: Biomechanics, fracture, internal fixation, locking plate, metacarpal

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Support was provided by Stryker Corporation, Freiburg, Germany, for implant and material costs. No benefits in any form have been received or will be received related directly or indirectly to the subject of this article.

PII: S0363-5023(10)00003-1

doi:10.1016/j.jhsa.2010.01.002

Journal of Hand Surgery
Volume 35, Issue 4 , Pages 597-603, April 2010